Skip to the content

FDA Medical Device Recalls (openFDA)

Integra LifeSciences Corp: Integra PAC1 Mod 2 Cables used with the Integra MPM, CAM01 Monitors

Recall number
Z-0944-2012
Recalling firm
Integra LifeSciences Corp
Product
Integra PAC1 Mod 2 Cables used with the Integra MPM, CAM01 Monitors
Status
Terminated
Initiated
2011-10-20
Posted
2012-02-03
Terminated
2013-06-18
Reason
Some of the connector cables associated with Integra's intracerebral pressure (ICP) monitors may cause temporary interference with the ICP waveform on the monitor display.
Root cause
Under Investigation by firm
Action
Integra sent a Urgent Product Recall Notification letter dated October 20, 2011, via FedEx, email, and fax to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The customer was instructed to examine their inventory at your earliest convenience. If you have an effected PAC1 cable, please complete the attached Recall Acknowledgement Form and return it to Integra Customer Service per the instructions on the form. An Integra Sales Representative will contact you to assist in the return and replacement of new cables at no cost. In the meanwhile, you may continue to use the cables until you have been contacted and replacement product has been provided. Distortion of the ICP wave form can be stopped or prevented by keeping the PACI catheter receptacle out of human contact. Please contact Integra Customer Service at 1-800-654-2873.
Quantity
221 Units
Distribution
Worldwide Distribution-- USA (nationwide ) .
Lot, serial or model codes
MPM, CAMO1, PAC1 :  MEF1100103;MED1102303; MEB1101602; MEF1100203; MEC1100602;MED1101003; MED1102203; MEE1102003; MED1102103; MEC1102302; MEB1101402; MEE1101003; MEF1100703; MEE1101903; MEF1101103; MEB1100803; MEC1103002; MEE1100903; MED1101703; MED1101203; MEC1102902; MEB1101103; MEC1102002; MEB1101502; MED1102203; MEF1100603; MEE1101203; MEC1100102; MEE1101503; MEE1100703; MEF1100403; MED1101303; MED1101403; MEF1100303; MEC1100702; MED1101103; MEC1102402; MEC1101602; MEF1100903; MEE1101303; MED1102003; MEC1103302; MEE1101803; MEE1100803; MEF1100503; MEC1102502; MEC1101702; MEE1101403; MEC1103402; MEF1100803; MEC1102802; MEC1101802;MEF1101003; MEC1103102; MEC1101902; MEC1102102
510(k) numbers
["K962928"]
PMA numbers
not on file
Product code
GWM
Firm FEI
1121308
Firm city
Plainsboro
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register