Breas Medical, Inc.: Breas Ventilator Trolley with the Vivo 50/60/65 and Vivo 45LS ventilator models Article no 007384

Recall number
Z-0943-2022
Recalling firm
Breas Medical, Inc.
Product
Breas Ventilator Trolley with the Vivo 50/60/65 and Vivo 45LS ventilator models Article no 007384
Status
Terminated
Initiated
2022-03-21
Posted
not on file
Terminated
2024-12-13
Reason
Ventilator mounting brackets for the Vivo 50/60/65 and Vivo 45LS ventilator models can become detached from the Breas Ventilator Trolley, this may cause the ventilator to fall from the trolley and cause injury to a person who is in the immediate proximity
Root cause
Process control
Action
Breas Medical issued recall information via telephone followed by Urgent Medical Device Recall letter to Distributors/End Users via email on 3/21/22 via telephone.. Letter state reason for recall, health risk and action to take: Breas Medical s solution is to provide Inspect and Repair Kits (REF 008591) to all those potentially affected by this notice. The Inspect and Repair Kits are used to inspect the Trolley to determine if remedial action is required, and to replace the defective part if needed. We will send you enough Inspect and Repair Kits (REF 008591)s for all trolleys in your possession. " We (the Customer) will identify Breas Ventilator Trolleys (REF 007384) delivered between June 11th, 2020, and December 3rd, 2021. " Breas will provide us with the corresponding number of Inspect and Repair Kits (REF 008591). " We will inspect the swivel assembly and complete the enclosed form to indicate whether the trolley contained the necessary washer and no further action is needed, or if the washer was missing and the swivel assembly was replaced by the new one. " We will scrap the replaced swivel assembly. e kindly ask you to acknowledge receipt of this Field Safety Notice together with your choice to remedy the issue using attached reply form Trolley Field Safety Notice Distributor/Customer Reply Form with Option Selection . Contact Information Email Christina.Joiner@breas.com Helpline +1 (855) 436-8724 Postal Address 16 Esquire Road, N. Billerica, MA 01862 Web Portal breas.com
Quantity
707 units
Distribution
US Nationwide
Lot, serial or model codes
Trolleys delivered between June 11th, 2020, and December 3rd, 2021. UDI: 07321820073849
510(k) numbers
not on file
PMA numbers
not on file
Product code
CBK
Firm FEI
3010817335
Firm city
North Billerica
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register