MicroPort Orthopedics Inc.: PROFEMUR(R) R RASP HANDLE, REF PPW38078, 1 EACH, NON-STERILE, MicroPort Orthopedics, Arlington, TN Instruments support the implantation of implants. Broach handles mate to the broach to allow for preparation of the femoral canal for the stem component to be implanted. This

Recall number
Z-0943-2017
Recalling firm
MicroPort Orthopedics Inc.
Product
PROFEMUR(R) R RASP HANDLE, REF PPW38078, 1 EACH, NON-STERILE, MicroPort Orthopedics, Arlington, TN Instruments support the implantation of implants. Broach handles mate to the broach to allow for preparation of the femoral canal for the stem component to be implanted. This
Status
Terminated
Initiated
2016-11-17
Posted
2017-01-03
Terminated
2018-08-18
Reason
The recall is being initiated because all broach handles in the affected lot had the same non-conformity with the connection feature. The broach handles may have machining issues that do not allow the handle to properly mate with an associated broach.
Root cause
Process control
Action
MicroPort Orthopedics sent an Voluntary Device Product Recall letter to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to complete and return the enclosed form even if they do not have any affected product to return. Return the completed form by fax to 1-901-451-6032 or by email to PostMarket@ortho.microport.com. Attempt #1 for acknowledgment will be made by email. For questions regarding this recall call 901-867-4318.
Quantity
20 units
Distribution
Worldwide Distribution to Spain, China, and Italy
Lot, serial or model codes
Lot Number US100122
510(k) numbers
["K012091"]
PMA numbers
not on file
Product code
LZO
Firm FEI
3010536692
Firm city
Arlington
Firm state
TN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register