FDA Medical Device Recalls (openFDA)
Stryker Medical Division of Stryker Corporation: Foot Rest Kit (pN 6252700-003) for Models 6251 and 6252 Stryker (EMS Stair -PRO), OE 2065939, 1739631, 2066066, 6066803 and 20666941 with optional , Stryker Medical, Portage, MI 49002. Intended to provide the patient greater comfort and a sense of security.
- Recall number
- Z-0943-2012
- Recalling firm
- Stryker Medical Division of Stryker Corporation
- Product
- Foot Rest Kit (pN 6252700-003) for Models 6251 and 6252 Stryker (EMS Stair -PRO), OE 2065939, 1739631, 2066066, 6066803 and 20666941 with optional , Stryker Medical, Portage, MI 49002. Intended to provide the patient greater comfort and a sense of security.
- Status
- Terminated
- Initiated
- 2011-10-31
- Posted
- 2012-02-06
- Terminated
- 2013-08-28
- Reason
- The Foot Rest Kit may have the wrong pivot spacers making the joint more susceptible to relaxation and prone to fatigue. There is a moderate risk of injury to the patient should the handles become detached during use.
- Root cause
- Mixed-up of materials/components
- Action
- Stryker Medical spoke with all customers who received these shipments prior to 10/31/2011 requesting information on kits affected and recalled foot rest units installed. December 5, 2011 Urgent Medical Device Recall Letters, with instructions on how to to either install a new Foot Rest Kit ( enclosed ) or dispose of any kits that were not installed were sent to both customers who could have installed recalled foot rest units. Customers needing help or assistance with installation are asked to call Stryker Technical Support a 1-800-327-0770. Customers were to complete and return a postage paid postcard confirming receipt. Customers with questions or concerns are asked to contact Kristin Dudek at 269-389-6927 M-F m 8 a.m. - 5 p.m. (EST)
- Quantity
- 3 units
- Distribution
- Nationwide Distribution -- MI & MA.
- Lot, serial or model codes
- Model 6251 and 6252, The Foot Rest Kit (pN 6252-700-003) Order numbers 2065939, 1739631, 2066066, 2066803, and 2066941.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- FPP
- Firm FEI
- 1831750
- Firm city
- Portage
- Firm state
- MI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28