FDA Medical Device Recalls (openFDA)
GE Healthcare, LLC: GE Discovery ST (16 Slice) System; Discovery ST (16 Slice) Systems with technical Reference Manual numbers 2351785-xxx and 2390935-xxx distributed from November 2001 to August 2010. Head and whole body X-ray computed tomography applications.
- Recall number
- Z-0943-2011
- Recalling firm
- GE Healthcare, LLC
- Product
- GE Discovery ST (16 Slice) System; Discovery ST (16 Slice) Systems with technical Reference Manual numbers 2351785-xxx and 2390935-xxx distributed from November 2001 to August 2010. Head and whole body X-ray computed tomography applications.
- Status
- Terminated
- Initiated
- 2010-10-27
- Posted
- 2011-02-07
- Terminated
- 2012-07-13
- Reason
- Although the system x-ray output is accurate for the technique factors given, and the dose display is also accurate, certain CTDI adjustment factors used for the manual calculation from the Technical Reference Manual (TRM) are incorrect. CTDI100 Aperture Adjustment Factors for 4x0.625 mode contain incorrect values. This issue may lead to inaccurate CTDI manual calculation. The results of the diffe
- Root cause
- Radiation Control for Health and Safety Act
- Action
- An Important Electronic Product Radiation Warning letter will be provided together with the TRM addendum informing the customers of the issue and related hazards. The letter also identified the affected product and gave instructions, which include referring to the TRM addendum, along with the product correction. This correction will be executed via a Field Modification Instruction IFMI 254281 released in November, 2010 with an estimated completion date of May, 2011. Questions or concerns should be directed to the call numbers listed or a local GE Healthcare Service Representative for countries not listed.
- Quantity
- 1450 units total in US
- Distribution
- Worldwide Distribution
- Lot, serial or model codes
- Technical Reference Manual numbers 2351785-xxx and 2390935-xxx.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- JAK
- Firm FEI
- 2126677
- Firm city
- Waukesha
- Firm state
- WI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28