FDA Medical Device Recalls (openFDA)
Draegar Medical Systems, Inc.: INFINITY DUAL HEMO MCable Pod
- Recall number
- Z-0942-2018
- Recalling firm
- Draegar Medical Systems, Inc.
- Product
- INFINITY DUAL HEMO MCable Pod
- Status
- Terminated
- Initiated
- 2017-10-04
- Posted
- not on file
- Terminated
- 2018-06-15
- Reason
- Dual Hemo MCable Pods with Revision Index (RI) 15 and 16, which are used with the Drger Infinity Acute Care System, may permit liquid ingress inconsistent with their IPX4 rating, which could potentially result in an incorrect measured value of invasive blood pressure.
- Root cause
- Nonconforming Material/Component
- Action
- In September of 2017 an Urgent medical Device Recall letter was issued to customers informing them of the recall and to request that inventory be checked to see if the product subject in the recall is on the premises. To do this, the customer will check the label for the revision index number which is located on the upper middle portion of label. RI number 15 and 16 are included in the recall. No other numbers are subject to recall. Affected product will receive a sleeve from the company to correct the issue. Questions or concerns can be directed to 1-800-437-2437 (Press 1, then press 32349) or contact Customer Care at 1-800-437-2437 (Press 4, then press 4 again)
- Quantity
- 2,760
- Distribution
- Nationally
- Lot, serial or model codes
- Have the affected revision indexes: 15 or 16 (found on the product label)
- 510(k) numbers
- ["K113798"]
- PMA numbers
- not on file
- Product code
- MHX
- Firm FEI
- 3005783425
- Firm city
- Andover
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28