Skip to the content

FDA Medical Device Recalls (openFDA)

Draegar Medical Systems, Inc.: INFINITY DUAL HEMO MCable Pod

Recall number
Z-0942-2018
Recalling firm
Draegar Medical Systems, Inc.
Product
INFINITY DUAL HEMO MCable Pod
Status
Terminated
Initiated
2017-10-04
Posted
not on file
Terminated
2018-06-15
Reason
Dual Hemo MCable Pods with Revision Index (RI) 15 and 16, which are used with the Drger Infinity Acute Care System, may permit liquid ingress inconsistent with their IPX4 rating, which could potentially result in an incorrect measured value of invasive blood pressure.
Root cause
Nonconforming Material/Component
Action
In September of 2017 an Urgent medical Device Recall letter was issued to customers informing them of the recall and to request that inventory be checked to see if the product subject in the recall is on the premises. To do this, the customer will check the label for the revision index number which is located on the upper middle portion of label. RI number 15 and 16 are included in the recall. No other numbers are subject to recall. Affected product will receive a sleeve from the company to correct the issue. Questions or concerns can be directed to 1-800-437-2437 (Press 1, then press 32349) or contact Customer Care at 1-800-437-2437 (Press 4, then press 4 again)
Quantity
2,760
Distribution
Nationally
Lot, serial or model codes
Have the affected revision indexes: 15 or 16 (found on the product label)
510(k) numbers
["K113798"]
PMA numbers
not on file
Product code
MHX
Firm FEI
3005783425
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register