FDA Medical Device Recalls (openFDA)
Edwards Lifesciences, Llc: Vigilance II Monitor Model: VIG2/VIG2E Software: versions 00.51, 00.60, 00.61, 00.62, 00.63, 00.64, 00.67, 00.69, 00.71, 00.73. Product Usage: The Vigilance II CCO/SvO2/CEDV Monitor is indicated for use in patients requiring monitoring of hemodynamic parameters, including cardiac output, oximetry and right ventricular, ejection, fraction, and end diastolic volume.
- Recall number
- Z-0942-2012
- Recalling firm
- Edwards Lifesciences, Llc
- Product
- Vigilance II Monitor Model: VIG2/VIG2E Software: versions 00.51, 00.60, 00.61, 00.62, 00.63, 00.64, 00.67, 00.69, 00.71, 00.73. Product Usage: The Vigilance II CCO/SvO2/CEDV Monitor is indicated for use in patients requiring monitoring of hemodynamic parameters, including cardiac output, oximetry and right ventricular, ejection, fraction, and end diastolic volume.
- Status
- Terminated
- Initiated
- 2008-09-12
- Posted
- 2012-02-02
- Terminated
- 2012-02-08
- Reason
- Edwards Lifesciences has made the decision to initiate a field corrective action for the Edwards Vigilance II Model: Software: versions 00.51, 00.60, 00.61, 00.62, 00.63, 00.64, 00.67, 00.69, 00.71, 00.73 due to the potential of inaccurate Injectate Cardiac Output (ICO) measurements.
- Root cause
- Device Design
- Action
- Edwards initiated two Urgent Product Notification - Action Required letters to all affected customers. The letters identified the affected product, problem, and actions to be taken. The letters instructed users to immediately discontinue use of in-line temperature probes for Intermittent Cardiac Output (ICO) measurements on the Vigilance II instrument and recommends use of the bath probe only (model 9850A) to measure the temperature of the injectate solution for ICO calculations, until software update have been completed. Customers were instructed to complete and return a software upgrade form to Edwards as record of completion. For questions call Edwards Technical Support department at 800-822-9837.
- Quantity
- 6934 instruments
- Distribution
- Worldwide Distribution - Nationwide Distribution (USA) - and the countries of: Argentina, Australia/ New Zealand, Brazil, Canada, China, Columbia, Ecuador, Europe, Indonesia, Japan, Korea, Macao, Malaysia, Mexico, Middle East, Peru, Puerto Rico, Singapore, Thailand, Taiwan and Venezuela
- Lot, serial or model codes
- Software: versions 00.51, 00.60, 00.61, 00.62, 00.63, 00.64, 00.67, 00.69, 00.71, 00.73. Serial numbers: VG000031 to VG007468
- 510(k) numbers
- ["K043103"]
- PMA numbers
- not on file
- Product code
- DXG
- Firm FEI
- 2015691
- Firm city
- Irvine
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28