FDA Medical Device Recalls (openFDA)
Stryker Medical Div. of Stryker Corporation: Stryker InTouch 1.0 Critical Care Bed (Model 2130) with the Integrated PositionPro mattress (Model 2920) option. The device is intended to be used as an Intensive Care Unit bed.
- Recall number
- Z-0942-2009
- Recalling firm
- Stryker Medical Div. of Stryker Corporation
- Product
- Stryker InTouch 1.0 Critical Care Bed (Model 2130) with the Integrated PositionPro mattress (Model 2920) option. The device is intended to be used as an Intensive Care Unit bed.
- Status
- Terminated
- Initiated
- 2009-01-05
- Posted
- 2009-02-25
- Terminated
- 2010-11-05
- Reason
- The touch screen on the footboard may display a "tilt over range" error, resulting in a precautionary stop of the bed motions which result in the bed appearing to have locked out some of its bed positioning features. Also, the "hold" button, used to interrupt an intended patient turn on beds with an integrated PositionPro mattress, may only function intermittently, resulting in an inability to ho
- Root cause
- Device Design
- Action
- Consignees were notified of the problems via letter dated January 5, 2009, advised to check their beds and to take them out of service if the foot board touchscreen displayed a "tilt over range" error, and that a Stryker Medical field service representative would contact them to perform software upgrades on the affected beds. Please direct questions to Stryker Medical at 1-269-324-6644.
- Quantity
- 1,288 of all models
- Distribution
- Worldwide Distribution including USA and countries of Canada, Japan and Switzerland.
- Lot, serial or model codes
- Serial Numbers: T02152, T02196, T02197, T02198, T02199, T02200, T02201, T02202, T02203, T02204, T02205, T02206, T02207, T02208, T02209, T02210, T02211, T02212, T02213, T02214, T02215, T02216, T02217, T02218, T02219, T02220, T02221, T02222, T02223, T02224, T02225, T02226, T02227, T02228, T02229, T02230, T02231, T02232, T02233, T02234, T02235, T02236, T02237, T01905, T01906, T01907, T01908, T01909, T01910, T01911, T01912, T01913, T01914, T01915, T01916, T01917, T01918, T01919, T01920, T01921, T01922, T01923, T01924, T01925, T01926, T01927, T01928, T01929, T01930, T01931, T01699, T01948, T01949, T01950, T01951, T01952, T01953, T01954, T01955, T01956, T01957, T01958, T01959, T01960, T01961, T01962, T01963, T01964, T01965, T01966, T01967, T01682, T01683, T01684, T01685, T01686, T01687, T01688, T01689, T01543, T01608, T01609, T01718, T01719, T01720, T02075, T02076, T02077, T02239, T02174, T01310, T01721, T01722, T01723, T02011, T02012, T02013, T02014, T02015, T02016, T02058, T02059, T02060, T02061, T02062, T02063, T02064, T02065, T01335, T01336, T01337, T01338, T01339, T01340, T01341, T01342, T01343, T01344, T01345, T01346, T01347, T01348, T01321, T01322, T01323, T01324, T01325, T01326, T01327, T01328, T01329, T01330, T01331, T01332, T01610, T01611, T01612, T01613, T02513, T02514, T02515, T01661, T01662, T01663, T01664, T01665, T01525, T01526, T01384, T01385, T01386, T01387, T01388, T02506, T02507, T02508, T02509, T02510, T02511, T02512, T02066, T02067, T02068, T01370, T01371, T01372, T01373, T01374, T01375, T01376, T01377, T01572, T01368, T01369, T02166, T02167, T02168, T02169, T02170, T02171, T02172, T02173, T01156, T01547, T01548, T01696, T01552, T01553, T01554, T01555, T02175, T02176, T02177, T02178, T02179, T02180, T02090, T02088, T01546, T01544, T02087, T01698, T02089, T01551, T01697, T01545, T01550 and T01695.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- FNL
- Firm FEI
- 1831750
- Firm city
- Portage
- Firm state
- MI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28