FDA Medical Device Recalls (openFDA)
Synthes USA HQ, Inc.: Large Combination Clamp/Large External Fixation System, Part Number 390.005 Product Usage: Synthes Large External Fixation is intended for use to provide treatment for long bone and pelvic fractures that require external fixation.
- Recall number
- Z-0941-2013
- Recalling firm
- Synthes USA HQ, Inc.
- Product
- Large Combination Clamp/Large External Fixation System, Part Number 390.005 Product Usage: Synthes Large External Fixation is intended for use to provide treatment for long bone and pelvic fractures that require external fixation.
- Status
- Terminated
- Initiated
- 2013-01-11
- Posted
- 2013-03-11
- Terminated
- 2015-08-31
- Reason
- The L1 hex nuts, used to tighten the Large Combination Clamp, were oversized and the 11 mm sockets and wrenches will not fit on the hex nuts.
- Root cause
- Nonconforming Material/Component
- Action
- Synthes sent an Urgent Notice: Medical Device Recall letter dated January 11, 2013 to all affected customers. The letter identified the affected product, problem, and actions to be taken. Customers were instructed to examine their inventory for the affected product(s), remove them from inventory and return them to Synthes. For questions call 610-719-5450 or email FieldAction@synthes.com
- Quantity
- 136
- Distribution
- USA Nationwide Distribution including the states of AL, AR, AK, CA, FL, ID, KS, LA, MD, MA, MO, NJ, NY, OH, PA, TN, TX, and VA.
- Lot, serial or model codes
- Lot Numbers 6602070, 6582333, 6581191, and 6581193
- 510(k) numbers
- K082650
- PMA numbers
- not on file
- Product code
- KTT
- Firm FEI
- 3005180112
- Firm city
- West Chester
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as