Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Healthcare Diagnostics, Inc.: Siemens ADVIA Chemistry Systems models 1200, 1650, 1800 and 2400 Product Usage: The ADVIA Chemistry Systems models 1200, 1650, 1800 and 2400 (also known as Clinical Chemistry Systems) are high throughput instruments which perform assays for general and specialty chemistries, methods for specific protein measurement, drugs-of-abuse testing and therapeutic drug monitoring.

Recall number
Z-0940-2014
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product
Siemens ADVIA Chemistry Systems models 1200, 1650, 1800 and 2400 Product Usage: The ADVIA Chemistry Systems models 1200, 1650, 1800 and 2400 (also known as Clinical Chemistry Systems) are high throughput instruments which perform assays for general and specialty chemistries, methods for specific protein measurement, drugs-of-abuse testing and therapeutic drug monitoring.
Status
Terminated
Initiated
2013-11-06
Posted
2014-02-09
Terminated
2015-08-10
Reason
The ADVIA Chemistry software systems using certain software versions allows the system to automatically run assay calibrations and quality control (QC).
Root cause
Software design
Action
Siemens sent an Urgent Medical Device Correct letter, dated November 6, 2013, to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to retain the recall notification letter with their laboratory records and forward to those who may have received the affected product.
Quantity
3759
Distribution
Worldwide Distribution and USA Nationwide.
Lot, serial or model codes
ADVIA 1200 V2.00 - 10386841 ADVIA 1200 V2.01 - 10469445 ADVIA 1650 V3.52 - 10318424 ADVIA 1650 V4.01 - 10285281 ADVIA 1800 V2.01 - 10639244 ADVIA 2400 V4.01 - 10639265
510(k) numbers
["K990346"]
PMA numbers
not on file
Product code
JJE
Firm FEI
2517506
Firm city
Newark
Firm state
DE
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register