FDA Medical Device Recalls (openFDA)
Symbios Medical Products, LLC: GOPump Elastometric infusion PumpKit with/ and without accesssories all containing Dual GOPump, part number 510076: One infusion pump comprising two, 150 mL elastomeric reservoirs, each of which connects to its own catheter. Liquid flow rate is labeled at 2 mL/hr per elastomer. The Symbios Disposable Infusion Pump Kit is a disposable, self-contained infusion system utilizing an inflatable elastomeric reservoir to mechanically provide percutaneous infusion of prescribed solutions at a pre-set rate for post-operative pain management
- Recall number
- Z-0940-2013
- Recalling firm
- Symbios Medical Products, LLC
- Product
- GOPump Elastometric infusion PumpKit with/ and without accesssories all containing Dual GOPump, part number 510076: One infusion pump comprising two, 150 mL elastomeric reservoirs, each of which connects to its own catheter. Liquid flow rate is labeled at 2 mL/hr per elastomer. The Symbios Disposable Infusion Pump Kit is a disposable, self-contained infusion system utilizing an inflatable elastomeric reservoir to mechanically provide percutaneous infusion of prescribed solutions at a pre-set rate for post-operative pain management
- Status
- Terminated
- Initiated
- 2013-02-25
- Posted
- 2013-03-13
- Terminated
- 2014-02-13
- Reason
- Received 2 complaints out of 458 distributed pumps in this lot, where flow restrictor bead became displaced, which were reported to permit fast flow of contents.
- Root cause
- Pending
- Action
- Symbios sent an Urgent Medical Device Recall letter dated February 25, 2013, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were instructed to follow the enclosed 'Steps for Voluntary Field Action' in their entirety. These steps contain information about: I) identifying the product to be returned, 2) obtaining a Returned Goods Authorization number, and 3) returning the recalled product. If product(s) is (are) located, please call 317-450-0145 from 7:00 a.In '0 5:00 p.m. EDT to receive a Returned Goods Authorization umber ("RGA"). Please indicate the RGA number on the outside of the package. Attached is a Verification Form. Customers were instructed to complete and return the form even if they do not have any product to return. Their local sales representative can assist them in completing this form. Distributors of the affected product were instructed to contact Symbios at the number provided and forward the recall notice to all their affected accounts. For questions regarding this recall call 317-225-4447.
- Quantity
- 458
- Distribution
- Nationwide Distribution including PA, CA, OH, AL, FL, WV, TN, NJ, NC, TX, CO, NY, TX, MO, MS, SC, and GA.
- Lot, serial or model codes
- Kit Part Number 510545-BP, lot 12-100983, 12-100984, 12-101096, 12-101125 Kit Part Number 510112-BP, lot 12-101110, 12-101103 Kit: 510401-BP lot 12-101045 Kit: 510644, lot 12-101028, 12-101097, 12-101130 GOPump Dual Infusion Pump Kit All Kits contain Pump Part Number: 510076 Lot Number: 12-100601
- 510(k) numbers
- ["K072921"]
- PMA numbers
- not on file
- Product code
- MEB
- Firm FEI
- 3005203102
- Firm city
- Indianapolis
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as