Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: Siemens MAGNETOM Aera/Skyra, model numbers 10432914 and 10432915. Nuclear magnetic resonance imaging system.

Recall number
Z-0940-2012
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Siemens MAGNETOM Aera/Skyra, model numbers 10432914 and 10432915. Nuclear magnetic resonance imaging system.
Status
Terminated
Initiated
2012-01-26
Posted
2012-02-02
Terminated
2014-01-31
Reason
During product monitoring firm has discovered that on a few systems the gradient cable connections did not meet their specifications after installation. Locally, the cable connections overheated, making servicing necessary. The potential existed for a cable connection to become overheated, leading to the emission of smoke and possible material damage. Firm preventively checked the gradient cabl
Root cause
Nonconforming Material/Component
Action
Siemens Medical Solutions sent a "CUSTOMER SAFETY ADVISORY NOTICE" letter dated November 24, 2010 to all affected customers. The letter described the product, problem, and actions to be taken. Contact the firm at 610-219-4834 for questions regarding this recall.
Quantity
128
Distribution
Nationwide Distribution-including the states of AK, AZ, CA, CT, DE, FL, GA, IL, IN, KS, MD, MA, MI, MN, MO, MT, NE, NH, NJ, NM, NY, NC, OH, OR, PA, SC, SD, TN, TX, UT, VA, WA, WV, and WI.
Lot, serial or model codes
Serial numbers 45115, 41187, 41157, 41116, 45072, 45081, 41089, 41215, 41144, 41201, 45148, 41164, 41088, 41245, 41168, 41152, 45140, 45152, 45143, 45195, 41222, 41150, 45129, 45118, 45119, 41134, 41138, 45182, 45213, 41167, 41100, 52024, 45196, 45197, 41180, 41130, 45088, 41189, 41104, 41246, 41176, 45138, 15084, 41071, 45120, 45107, 41271, 45135, 45141, 41264, 45110, 41279, 41231, 41284, 41235, 45221, 45223, 41276, 41278, 41137, 41225, 41131, 41141, 41233, 45185, 41236, 41238, 45149, 41191, 45074, 45106, 45006, 45201, 41133, 41265, 41188, 45220, 41140, 41079, 45160, 45167, 41194, 45155, 45130, 41241, 41170, 45147, 41006, 45219, 45085, 45150, 45031, 41162, 45121, 45134, 45099, 45038, 45079, 41109, 41156, 45116, 41249, 45128, 45078, 45214, 41103, 45229, 41234, 45181, 52025, 52021, 45105, 45132, 41149, 45228, 45060, 45080, 45208, 41083, 45230, 45205, 41097, 41181, 45093, 45199, 45184, 45071, and 45111.
510(k) numbers
["K101347"]
PMA numbers
not on file
Product code
LNH
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register