Skip to the content

FDA Medical Device Recalls (openFDA)

Maquet Cardiovascular, LLC: Maquet Heartstring II Proximal Seal System (HS-1045) manufactured by Maquet Cardiovascular, Wayne, NJ Intended use: medical device for use by cardiac surgeons during CABG procedures to maintain hemostasis and to facilitate the completion of a proximal anastomosis without application of an aortic clamp.

Recall number
Z-0940-2011
Recalling firm
Maquet Cardiovascular, LLC
Product
Maquet Heartstring II Proximal Seal System (HS-1045) manufactured by Maquet Cardiovascular, Wayne, NJ Intended use: medical device for use by cardiac surgeons during CABG procedures to maintain hemostasis and to facilitate the completion of a proximal anastomosis without application of an aortic clamp.
Status
Terminated
Initiated
2010-05-05
Posted
2011-02-01
Terminated
2011-06-21
Reason
Deployment tube may detach during use due to insufficient adhesive, which may result in product failure to operate.
Root cause
Process design
Action
The firm, MAQUET Getinge Group, sent an "URGENT DEVICE REMOVAL IMMEDIATE ACTION REQUIRED" letter dated May 5, 2010, to all customers via Federal Express. The letter described the product, problem and actions to be taken by the customer. The customers were instructed to immediately examine their stocks to determine if they have any of the affected product, if found, discontinue dispensing (distributing) the lot and complete the Field Action Response Form; immediately contact accounts if they acted as a distributor, and acknowledge receipt of this notification whether or not they have any affected products by completing the Field Action Response Form via fax to (U.S.): 1-888-899-2874. If you have any additional questions, please contact your local MAQUET Cardiovascular sales representative or our Customer Service at 1-888-880-2874.
Quantity
1630
Distribution
Worldwide distribution: USA and countries including: Japan, Mexico, Germany, Switzerland, Israel
Lot, serial or model codes
Lot numbers 12895788, 12946718, 25002866, 25004647, 13029141, 25005045, 25005622
510(k) numbers
["K022880"]
PMA numbers
not on file
Product code
DXC
Firm FEI
2242352
Firm city
Wayne
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register