Skip to the content

FDA Medical Device Recalls (openFDA)

Instrumentation Laboratory: HemosIL SynthAFax. Partial Thromboplastin Time Tests.

Recall number
Z-0938-2026
Recalling firm
Instrumentation Laboratory
Product
HemosIL SynthAFax. Partial Thromboplastin Time Tests.
Status
Open, Classified
Initiated
2025-11-18
Posted
2025-12-17
Terminated
not on file
Reason
Recalled lots were manufactured with double the amount of preservative concentration.
Root cause
Process control
Action
An URGENT MEDICAL DEVICE RECALL notification, dated 11/18/25, was sent to consignees. Consignees are to contact Client Services at 1-800-955-9525 to arrange conversion to replacement lot N0956748, and in the meantime continue use of affected lots with the QC lot reassignments provided and run QC per labeled instructions for use. Once the replacement lot is received, consignees are to complete the cross-over study per laboratory procedures and immediately destroy affected devices, documenting destruction on the provided Mandatory Urgent Medical Device Recall Tracking Form which should be returned to WERFEN by fax at 781-861-4207 or by email to ra-usa.na@werfen.com. All laboratory staff are to be notified of this recall and a copy of the notice posted near IL Coagulation Systems. Consignees with any questions are to contact Client Services at 1-800-955-9525.
Quantity
4,506 units
Distribution
Domestic: CA, CO, FL, NY; Foreign: AR, AT, CA, CZ, DE, DK, ES, FR, GB, IE, IL, IN, IT, JP, LK, NL, PL, PT, RO, RS, TH, VN.
Lot, serial or model codes
Part No. 0020007400; UDI: 08426950087649; Lot No. (Exp. Date) N0542231(2026-06-30), N0250088(2027-04-30, N0452249 (2027-06-30).
510(k) numbers
["K955638"]
PMA numbers
not on file
Product code
GFO
Firm FEI
1217183
Firm city
Bedford
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register