FDA Medical Device Recalls (openFDA)

Graphic Controls Acquisition Corporation: CLARAVUE Pre-wired Electrode - IP SET 12 RT ADULT-US Model: 32028779 50602-US

Recall number
Z-0938-2022
Recalling firm
Graphic Controls Acquisition Corporation
Product
CLARAVUE Pre-wired Electrode - IP SET 12 RT ADULT-US Model: 32028779 50602-US
Status
Terminated
Initiated
2022-03-15
Posted
not on file
Terminated
2024-07-12
Reason
Instructions for Use (IFU) (Instructions for Use) indicate compliance with ANSI/AAMI EC53:2013/(R)2020 ECG trunk cables and patient leadwires. Section 5.3.9 of the standard requires the device to withstand, without breakdown , an application of, 5,000 V d.c. when used with a defibrillator at a voltage above its maximum threshold, Claravue does not meet the specification and delivers an inefficient electrical charge which may result in an ineffective defibrillation.
Root cause
Device Design
Action
Graphic Controls (d/b/a Nissha Medical Technologies/Vermed/Biomedical Innovations) issued Urgent Field Correction letter via email on 3/4/22. Letter states reason for recall, health risk and action to take: Section 5.1.5 of the standard provides that trunk cables and patient leadwires may be permanently marked to indicate they are not defibrillation proof. Immediately examine your inventory and complete the provided form Claravue IP Set Correction Notice Acknowledgement Form and return to claravue@nisshamedical.com If you have product in inventory our team will arrange for it to be returned and exchanged for product labeled correctly once available. In addition, if you may have further distributed this product, please identify your customer and notify them at once of this product correction. Your notification may be enhanced by including a copy of this correction letter. Please complete and return the attached form via e-mail to jpotter@nisshamedical.com as soon as possible. If you have any questions about this notice, please contact: Ms. Jessica Potter, Vice President Quality and Regulatory Nissha Medical Technologies, 400 Exchange Street, Buffalo, NY 14204, USA Phone: (716) 849-6367 E-Mail: jpotter@nisshamedical.com
Quantity
8265 units
Distribution
US Nationwide distribution in the States of AR, CA, CO, FL, IN, KS, MA, MN, NY, OK, VT.
Lot, serial or model codes
All lots UDI: 03700506307326
510(k) numbers
["K101685"]
PMA numbers
not on file
Product code
DRX
Firm FEI
1317188
Firm city
Buffalo
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register