SEASPINE ORTHOPEDICS CORPORATION: Orthofix NorthStar OCT Navigation Surgical Technique Guide, and instruments: Navigation Driver, REF: PC2-400012; Navigation Straight Bone Probe, REF: PC2-400013; Navigation, 3.5mm Drill, REF: PC2-400135; Navigation, 4.0mm Drill, REF: PC2-400140; Navigation, 4.5mm Drill, PC2-400145; Navigation, 5.5mm Drill, REF: PC2-400155; Navigation, 3.5mm Tap, REF: PC2-400235; Navigation, 4.0mm Tap, REF: PC2-400240; Navigation, 4.5mm Tap, REF: PC2-400245; Navigation, 4.5mm Dual Lead Tap, REF: PC2-400345; Navigation, 5.5mm Dual Lead Tap, REF: PC2-400355

Recall number
Z-0937-2026
Recalling firm
SEASPINE ORTHOPEDICS CORPORATION
Product
Orthofix NorthStar OCT Navigation Surgical Technique Guide, and instruments: Navigation Driver, REF: PC2-400012; Navigation Straight Bone Probe, REF: PC2-400013; Navigation, 3.5mm Drill, REF: PC2-400135; Navigation, 4.0mm Drill, REF: PC2-400140; Navigation, 4.5mm Drill, PC2-400145; Navigation, 5.5mm Drill, REF: PC2-400155; Navigation, 3.5mm Tap, REF: PC2-400235; Navigation, 4.0mm Tap, REF: PC2-400240; Navigation, 4.5mm Tap, REF: PC2-400245; Navigation, 4.5mm Dual Lead Tap, REF: PC2-400345; Navigation, 5.5mm Dual Lead Tap, REF: PC2-400355
Status
Open, Classified
Initiated
2025-11-17
Posted
2025-12-17
Terminated
not on file
Reason
The NorthStar OCT Navigation Surgical Technique Guide (STG) incorrectly identified both Medtronic s Solera and Medtronic s Infinity system toolcards as compatible with the NorthStar OCT Navigation Instruments. However, the NorthStar OCT Navigation Instruments are compatible for use with Medtronic s Solera toolcards only.
Root cause
Labeling design
Action
On 10/27/2025, device correction notices were emailed to customers who were informed of the following: The NorthStar OCT Navigation Surgical Technique Guide (STG) incorrectly identified both Medtronic's Solera and Medtronics Infinity system StealthStation tool cards as compatible with the NorthStar OCT Navigation Instruments; however, the NorthStar OCT Navigation Instruments are indicated for use with Medtronic's Solera tool cards only. 1) Discontinue use of Medtronic's Infinity tool cards with NorthStar OCT Navigation instruments. 2) Discard affected version of the Surgical Technique Guide 3) This notice should be shared within your organization and forwarded to any organization where affected Instruments and guide is utilized. 4) Review updated Surgical Technique Guide D0003926 Revision DA. This notice should be shared within your organization and forwarded to any organization where affected instruments are utilized. Any product complaints should be reported to Orthofix at complaints@orthofix.com
Quantity
310
Distribution
US: CT, TX, NV, CO, CA, WA, NJ, OR, IN, MI, FL, PA, IL, VA, AL, RI, MO, AZ, MS, SC, UT, NY, OK, SD. OUS: Australia, Mexico, Chile
Lot, serial or model codes
Surgical Technique Guide revisions: D0003926A, D0003926B, D0003926C, D0003926D. REF/UDI-DI/Lot: PC2-400012/10889981260033/MK81527A, MK81527ARR, MK81527ARRR, MK81574AR, MK81574AR1, MK93487A, MK93488ARR, MK93489A, MK93490A, MK93490ARR, MK93491AR, MK93491AR1, MK943488A, MK95882A, MK95882ARR, ML52083B, ML53136B, ML53188B, ML69777B, ML69778B, ML71257B, MM1001498D, MM1001499D, MM1001748D, MM1001749D, MM1001813D, MM1001902D, MM1001903D, MM1002025D, MM1002316D, MM1002328D, MM1002344D; PC2-400013/10889981260040/AL657979C, AL658549C, AL658549CRRR, AL658641C, AL661470C, AL671917D, AL683456F, AL702785F; PC2-400135/10889981260057/BK104899A, BK104899AR, BK104899ARR, BK104899ARRR, BK104899ARRR4, BK106822A, CR49020D, CR49023D, CR49060D, CR51550E, CR51972D, CR56472E; PC2-400140/10889981260064/BK104900A, BK104900AR, BK104900ARRR, BK106825A, BK107806A, BK107806ARRR, BK108366A, BK108366ARRR, CR49822D, CR50158D, CR51973D, CR57547E, CR59411E; PC2-400145/10889981260071/BK104901A, BK104901ARRR, BK106827A, CR49021C, CR49061C, CR51974C; PC2-400155/10889981260088/BK104902A, BK104902ARRR, BK106829A, CR49062C, CR51551C, CR51975C, CR57550E, CR59413E; PC2-400235/10889981260095/BK104903A, BK104903ARRR, BK106831A, BK106831ARRR, BK131159C, CR49620C, CR52371D, CR54729D, CR58013D, CR59414D; PC2-400240/10889981260101/BK104904A, BK104904ARRR, BK106833A, BK131160C, CR49621C, CR52372D, CR59415D; PC2-400245/10889981275495/no lots released; PC2-400345/10889981260118/BK104905A, BK104905ARRR, BK106834A, BK106834ARRR, BK131161C, CR49622C, CR52373D, CR54208D; PC2-400355/10889981260125/BK104906A, BK106835A, CR51764C, CR52853D
510(k) numbers
["K172517"]
PMA numbers
not on file
Product code
OLO
Firm FEI
3012120772
Firm city
Carlsbad
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register