Skip to the content

FDA Medical Device Recalls (openFDA)

Vygon Corporation: Churchill Medical Systems, Inc. A Vygon Company - All sizes of CMS-CT6T225 6F Triple Lumen CT PICC with M.S.T. Components, Peripherally Inserted Central Catheter.

Recall number
Z-0937-2013
Recalling firm
Vygon Corporation
Product
Churchill Medical Systems, Inc. A Vygon Company - All sizes of CMS-CT6T225 6F Triple Lumen CT PICC with M.S.T. Components, Peripherally Inserted Central Catheter.
Status
Terminated
Initiated
2012-12-17
Posted
2013-03-11
Terminated
2014-01-14
Reason
Product was packaged with the incorrect introducer needle. The kit label indicates a "safety" needle is included; the kits were packaged with a non-safety needle.
Root cause
Component change control
Action
Churchill Medical Systems sent a letter to their consignees on December 19, 2012.
Quantity
210
Distribution
Product was distributed to 4 hospitals (under one hospital network) within the State of Pennsylvania
Lot, serial or model codes
Lots numbers MBKQ640, exp 05/2015; MBKZ720, exp 05/2015; MBLJ290, exp 06/2015; MBLR100, exp 07/2015; MBLZ350, exp 09/2015; and MBMG250, exp 10/2015.
510(k) numbers
["K081904"]
PMA numbers
not on file
Product code
LJS
Firm FEI
2245270
Firm city
Lansdale
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register