Zoll Lifecor Corporation: Electrode Belt, component of the LifeVest Wearable Defibrillator, Product number 10A0889-A01. Specifically the 3 Therapy Electrode Pads attached to the Electrode Belt.

Recall number
Z-0936-2007
Recalling firm
Zoll Lifecor Corporation
Product
Electrode Belt, component of the LifeVest Wearable Defibrillator, Product number 10A0889-A01. Specifically the 3 Therapy Electrode Pads attached to the Electrode Belt.
Status
Terminated
Initiated
2007-04-27
Posted
2007-06-15
Terminated
2008-03-06
Reason
The electrodes do not properly release conductive gel
Root cause
Other
Action
Patients currently using the product should continue to do so while the product removal is underway. Product replacements will begin with the belts currently in the possession of the sales reps. By replacing these first any new patients will be receiving new or refurbished belts with new Therapy Electrodes that have not been exposed to methanol. As patients complete their use of the LifeVest and return their device to the firm the belt's refurbishment process will include replacement of all 3 Therapy Electrodes. The firm expects this process to be completed in approximately 3 months based on the average use of time and current production capabilities. The process may take an additional month for devices distributed overseas. Physicians with active patients were notified via letter sent certified mail on 5/3/07. Long term use patients (patients using the device longer than 3 months) will be notified of the product removal via telephone and follow up letter. Each long term patient will be provided with a replacement belt. They will be instructed to return their existing belt in the same shipping box. Due to the low residual risk no notification is planned for short term use patients. Prescribing physicians have the option to notify their patients if they choose to do so.
Quantity
944 belts
Distribution
Worldwide: The product was shipped to patients in AZ, CA, CT, FL, GA, IL, IN, KY, LA, MA, MI, MN, MO, MT, NC, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WI, and WV. The product was also shipped to Germany, United Kingdom, Italy, Spain, France, and Denmark.
Lot, serial or model codes
Product number 10A0889-A01. Serial numbers with active patients: 79000004, 79000055, 79000064, 79000073, 79000106, 79000157, 79000181, 79000205, 79000208, 79000217, 79000226, 79000238, 79000244, 79000259, 79000274, 79000277, 79000286, 79000289, 79000298, 79000328, 79000346, 79000358, 79000364, 79000367, 79000373, 79000376, 79000379, 79000388, 79000403, 79000409, 79000445, 79000454, 79000457, 79000466, 79000487, 79000490, 79000493, 79000508, 79000511, 79000529, 79000532, 79000547, 79000556, 79000562, 79000574, 79000577, 79000595, 79000598, 79000601, 79000604, 79000616, 79000631, 79000634, 79000637, 79000640, 79000652, 79000661, 79000670, 79000686, 79000685, 79000694, 79000697, 79000715, 79000727, 79000730, 79000739, 79000754, 79000760, 79000769, 79000790, 79000796, 79000799, 79000802, 79000811, 79000823, 79000829, 79000832, 79000835, 79000838, 79000841, 79000844, 79000847, 79000853, 79000856, 79000886, 79000889, 79000901, 79000904, 79000925, 79000931, 79000934, 79000937, 79000943, 79000958, 79000973, 79000979, 79000982, 79000994, 79001003, 79001018, 79001021, 79001024, 79001027, 79001030, 79001033, 79001036, 79001042, 79001048, 79001054, 79001060, 79001066, 79001078, 79001081, 79001093, 79001099, 79001102, 79001111, 79001117, 79001147, 79001150, 79001153, 79001156, 79001162, 79001198, 79001201, 79001207, 79001216, 79001219, 79001228, 79001237, 79001243, 79001255, 79001273, 79001276, 79001288, 79001300, 79001318, 79001339, 79001354, 79001357, 79001363, 79001369, 79001372, 79001375, 79001378,79001384, 79001387, 79001399, 79001402, 79001405, 79001408, 79001417, 79001423, 79001429, 79001441, 79001444, 79001456, 79001459, 79001462, 79001468, 79001486, 79001492, 79001498, 79001501, 79001504, 79001522, 79001525, 79001528, 79001534, 79001537, 79001540, 79001558, 79001564, 79001570, 79001573, 79001582, 79001585, 79001588, 79001594, 79001597, 79001603, 79001630 79001639, 79001642, 79001648, 79001654, 79001657, 79001660, 79001666, 79001672, 79001675, 79001681, 79001690, 79001693, 79001699, 79001714, 79001717, 79001732, 79001738, 79001741, 79001744, 79001750, 79001753, 79001756, 79001759, 79001762, 79001765, 79001771, 79001777, 79001780, 79001786, 79001795, 79001798, 79001801, 79001804, 79001810, 79001822, 79001834, 79001837, 79001840, 79001846, 79001849, 79001852, 79001867, 79001879, 79001882, 79001906, 79001912, 79001918, 79001924, 79001939, 79001942, 79001945, 79001954, 79001957, 79001963, 79001969, 79001984, 79001996, 79002002, 79002005, 79002011, 79002017, 79002026, 79002041, 79002044, 79002047, 79002053, 79002056, 79002059, 79002068, 79002071, 79002074, 79002089, 79002110, 79002116, 79002119, 79002122, 79002128, 79002140, 79002152, 79002155, 79002164, 79002170, 79002176, 79002179, 79002182, 79002185, 79002191, 79002194, 79002200, 79002206, 79002215, 79002218, 79002221, 79002224, 79002227, 79002233, 79002236, 79002248, 79002251, 79002263, 79002275, 79002278, 79002287, 79002290, 79002293, 79002305, 79002308, 79002320, 79002329, 79002332, 79002338, 79002347, 79002359, 79002368, 79002377, 79002392, 79002404, 79002413, 79002416, 79002422, 79002425, 79002428, 79002440, 79002446, 79002449, 79002452, 79002455, 79002458, 79002479, 79002482, 79002494, 79002497, 79002509, 79002512, 79002521, 79002524, 79002530, 79002533, 79002539, 79002554, 79002560, 79002563, 79002566, 79002578, 79002587, 79002593, 79002602, 79002605, 79002608, 79002617, 79002620, 79002623, 79002629, 79002635, 79002644, 79002647, 79002650, 79002659, 79002665, 79002671, 79002674, 79002677, 79002698, 79002701, 79002710, 79002728, 79002740, 79002749, 79002761, 79002764, 79002767, 79002770, 79002776, 79002779, 79002782, 79002785, 79002788, 79002794, 79002800, 79002806, 79002818, 79002827, 79002833, 79002842, 79002848, 79002860, 79002863, 79002866, 79002869, 79002875, 79002878, 79002881, 79002884, 79002896, 79002899, 79002902, 79002911, 79002917, 79002932, 79002938, 79002941, 79002950, 79002962, 79002965, 79002968, 79002971, 79002974, 79002977, 79002980, 79002983, 79002989, 79002992, 79003007, 79003010, 79003016, 79003019, 79003028, 79003031, 79003034, 79003037, 79003049, 79003052, 79003055, 79003058, 79003067, 79003070, 79003079, 79003082, 79003088, 79003097, 79003103, 79003118, 79003121, 79003130, 79003133, 79003136, 79003142, 79003151, 79003157, 79003166, 79003172, 79003175, 79003178, 79003190, 79003196, 79003199, 79003205, 79003211, 79003223, 79003229, 79003235, 79003244, 79003262, 79003265, 79003274, 79003277, 79003295, 79003316, 79003325, 79003328, 79003331, 79003346, 79003349, 79003373, 79003382, 79003400, 79003415, 79003445, 79003454, 79003457, 79003460, 79003463, 79003469, 79003472, 79003478, 79003493, 79003508, 79003511, 79003520, 79003526, 79003529, 79003538, 79003544, 79003550, 79003553, 79003556, 79003568, 79003583, 79003589, 79003592, 79003598, 79003601, 79003604, 79003607, 79003613, 79003616, 79003628, 79003631, 79003634, 79003637, 79003640, 79003646, 79003649, 79003652, 79003658, 79003661, 79003664, 79003670, 79003673, 79003697, 79003700, 79003703, 79003706, 79003715, 79003718, 79003721, 79003727, 79003730, 79003748, 79003757, 79003760, 79003763, 79003766, 79003769, 79003778, 79003784, 79003793, 79003796, 79003808, 79003823, 79003835, 79003838, 79003841, 79003844, 79003847, 79003850, 79003856, 79003862, 79003868, 79003871, 79003877, 79003889, 79003916, 79003919, 79003922, 79003949, 79003958, 79003961, 79003967, 79003976, 79003985, 79003988, 79004000, 79004003, 79004012, 79004021, 79004027, 79004042, 79004066, 79004078, 79004087, 79004096, 79004108, 79004117, 79004120, 79004126, 79004129, 79004135, 79004141, 79004144, 79004153, 79004162, 79004165, 79004210, and 79004216. Foreign serial numbers with active patients: 79000118, 79000724, 79000736, 79000343, 79000664, 79001075, 79001213, 79000310, 79000394, 79000139, 79000187, 79001231, 79001600, 79000256, 79000370, 79000463, 79000514, 79000592, 79000703, 79000745, 79000766, 79000808, 79001090, 79001171, 79001177, 79001315, 79001336, 79001345, 79001351, 79001390, 79001615, 79001825, 79001870, 79001900, 79001960, 79001966, 79002008, 79002062, 79002149, 79002203, 79002230, 79002272, 79002398, 79002464, 79002488, 79002491, 79002500, 79002545, 79002551, 79002584, 79002590, 79002611, 79002614, 79002680, 79002812, 79002953, 79003013, 79003022, 79003025, 79003094, 79003100, 79003124, 79003127, 79003139, 79003232, 79003247, 79003319, and 79003355. Serial numbers still under the firm''s control: 79000058, 79000190, 79000196, 79000211, 79000214, 79000241, 79000247, 79000253, 79000283, 79000325, 79000334, 79000340, 79000352, 79000391, 79000427, 79000439, 79000448, 79000475, 79000541, 79000544, 79000589, 79000643, 79000667, 79000712, 79000721, 79000772, 79000781, 79000784, 79000793, 79000814, 79000826, 79000850, 79000859, 79000862, 79000865, 79000874, 79000880, 79000895, 79000910, 79000913, 79000952, 79000961, 79000964, 79000985, 79000988, 79001012, 79001063, 79001072, 79001105, 79001120, 79001123, 79001138, 79001180, 79001225, 79001240, 79001252, 79001282, 79001291, 79001297, 79001303, 79001309, 79001312, 79001321, 79001324, 79001333, 79001360, 79001396, 79001420, 79001471, 79001489, 79001507, 79001546, 79001579, 79001591, 79001612, 79001621, 79001702, 79001723, 79001729, 79001792, 79001819, 79001828, 79001855, 79001861, 79001888, 79001903, 79001909, 79001927, 79001933, 79001951, 79001975, 79001981, 79001990, 79002032, 79002035, 79002038, 79002050, 79002065, 79002092, 79002095, 79002101, 79002125, 79002131, 79002134, 79002143, 79002146, 79002158, 79002173, 79002209, 79002245, 79002257, 79002269, 79002281, 79002302, 79002311, 79002314, 79002326, 79002350, 79002356, 79002395, 79002401, 79002410, 79002419, 79002434, 79002461, 79002470, 79002476, 79002569, 79002572, 79002575, 79002581, 79002596, 79002599, 79002632, 79002662, 79002683, 79002686, 79002692, 79002704, 79002713, 79002734, 79002743, 79002746, 79002773, 79002791, 79002803, 79002845, 79002851, 79002857, 79002893, 79002908, 79002923, 79002929, 79002935, 79002956, 79003019, 79003046, 79003085, 79003112, 79003148, 79003154, 79003160, 79003163, 79003169, 79003184, 79003187, 79003193, 79003214, 79003217, 79003226, 79003238, 79003241, 79003250, 79003253, 79003298, 79003301, 79003313, 79003322, 79003334, 79003340, 79003352, 79003379, 79003475, 79003481, 79003490, 79003496, 79003514, 79003541, 79003547, 79003571, 79003574, 79003577, 79003586, 79003643, 79003709, 79003724, 79003742, 79003751, 79003787, 79003790, 79003805, 79003814, 79003817, 79003832, 79003859, 79003865, 79003874, 79003892, 79003895, 79003904, 79003913, 79003925, 79003937, 79003940, 79003943, 79003952, 79003994, 79004006, 79004009, 79004030, 79004033, 79004036, 79004039, 79004045, 79004048, 79004063, 79004069, 79004072, 79004090, 79004093, 79004099, 79004102, 79004105, 79004111, 79004114, 79004132, 79004138, 79004147, 79004156, 79004159, 79004183, 79004186, 79004201, 79004204, 79004207, and 79004240. Foreign serial numbers still under the firm''s control: 79000361, 79000817, 79001132, 79001915, 79000499, 79000868, 79000898, 79001009, 79001414, 79001432, 79002284, 79003004, 79001258, 79001264, 79001279, 79001234, 79001249, 79000187, 79000559, 79001381, 79000265, 79000292, 79000349, 79000406, 79000424, 79000469, 79000520, 79000565, 79000625, 79000649, 79000673, 79000691, 79000877, 79001039, 79001087, 79001435, 79001474, 79001495, 79001513, 79001567, 79001606, 79001609, 79001774, 79001885, 79001930, 79001948, 79002020, 79002299, 79002365, 79002443, 79002503, 79002506, 79002668, 79002689, 79002755, 79002986, 32687900, 79003292, 79003310, 79003337, 79003358, 79003364, 79003403, 79003406, 79003409, and 79003412.
510(k) numbers
not on file
PMA numbers
["P010030"]
Product code
MVK
Firm FEI
3008642652
Firm city
Pittsburgh
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register