GE Healthcare, LLC: GE Healthcare, Innova 2100 IQ, Innova 3100 IQ, Innova 4100 IQ. Cardiovascular X-ray imaging systems. For use in generating fluoroscopic images of human anatomy.

Recall number
Z-0934-2013
Recalling firm
GE Healthcare, LLC
Product
GE Healthcare, Innova 2100 IQ, Innova 3100 IQ, Innova 4100 IQ. Cardiovascular X-ray imaging systems. For use in generating fluoroscopic images of human anatomy.
Status
Terminated
Initiated
2011-02-04
Posted
2013-03-11
Terminated
2014-01-27
Reason
Exposed high voltage wires inside the cabinet . On a cable located in the system cabinet, two of the shrink tubings covering the 400V live cable leave an area potentially unprotected.
Root cause
Nonconforming Material/Component
Action
No consignee letter sent out. GE Healthcare's recall strategy was to replace the impacted cable on all affected Innova 2100 IQ, Innova 3100 IQ, Innova 4100 IQ systems that were distributed. At this time, deployment of this recall is considered complete as of 9/6/2011.
Quantity
48
Distribution
Worldwide Distribution-USA including the states of FL, IL, IN, NJ, NC, OK, TX, and WI, and the countries of UNITED KINGDOM, TUNISIA, SERBIA, RUSSIAN FEDERATION, NORWAY, NIGERIA, LITHUANIA, LEBANON, KOREA, REPUBLIC OF KOREA, ITALY, INDONESIA, INDIA, FRANCE, FINLAND, COLOMBIA, CHINA, and BRAZIL.
Lot, serial or model codes
SERIAL NUMBER 592043BU4 592676BU1 593459BU1 593465BU8 592458BU4 592673BU8 592682BU9 593059BU9 593051BU6 593475BU7 593476BU5 592049BU1 593468BU2 593054BU0 592047BU5 592042BU6 592046BU7 592459BU2 593048BU2 593056BU5 593049BU0 593063BU1 593062BU3 593464BU1 594762BU7 593473BU2 592451BU9 593460BU9 597954BU7 592456BU8 592674BU6 593052BU4 593058BU1 590867BU8 592050BU9 593064BU9 592460BU0 592457BU6 592680BU3 592455BU0 593462BU5 592678BU7 594763BU5 593055BU7 592675BU3 592452BU7 593463BU3 594385BU7
510(k) numbers
["K050489","K052412","K061163","K092004"]
PMA numbers
not on file
Product code
MQB
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register