Skip to the content

FDA Medical Device Recalls (openFDA)

Smith & Nephew, Inc.: VERSABOND AB 40 GRAMS FORMULATION 2, antibiotic infused bone cement, Product Number: 71271440 - Product Usage: VERSABOND is a medium viscosity bone cement which can be used for hip, knee, and shoulder implant fixation. VERSABOND AB includes the antibiotic- gentamycin.

Recall number
Z-0933-2021
Recalling firm
Smith & Nephew, Inc.
Product
VERSABOND AB 40 GRAMS FORMULATION 2, antibiotic infused bone cement, Product Number: 71271440 - Product Usage: VERSABOND is a medium viscosity bone cement which can be used for hip, knee, and shoulder implant fixation. VERSABOND AB includes the antibiotic- gentamycin.
Status
Terminated
Initiated
2020-12-18
Posted
not on file
Terminated
2023-05-01
Reason
Potential heat and humidity exposure. The affected lot contained an atypical sandy, flaky or grainy consistency when mixing.
Root cause
Storage
Action
A Recall notification letter titled, "Urgent Medical Device Recall Notice" was sent to consignees on 12/18/2020 via email and FedEx. The letter instructs the consignee to follow the instructions on the attached response form. The response form provides the following instructions: "1. Please inspect your inventory and locate any devices from the listed product and batch numbers enclosed with this Urgent Medical Device Recall Notice, and quarantine them immediately. a. If you are a sales rep or distributor, you must notify your customers of the field action and ensure that the required actions are complete. b. Include the customers account number and name in the table below. 2. If you have no product to return, please put an X in the appropriate location below. 3. If you have product to return, please list the item, batches and quantities that you are returning in the appropriate boxes below. 4. Complete the remainder of the form, sign and send to FieldActions@smith-nephew.com. Please Note even if you have no product to return, this form must be completed, signed and returned. 5. Once the form is received by Smith & Nephew, you will be sent a Return Authorization (RA) number. If you have any questions or concerns regarding this recall please contact FieldActions@smith-nephew.com."
Quantity
5190 devices
Distribution
Worldwide distribution - US Nationwide distribution in the state of OH and the countries of Dubai, Canada, Korea, Chile, Puerto Rico, Panama, and Costa Rica.
Lot, serial or model codes
Product Number: 71271440; Batch Number: 18LC07540
510(k) numbers
["K022688"]
PMA numbers
not on file
Product code
LOD
Firm FEI
1020279
Firm city
Memphis
Firm state
TN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register