Skip to the content

FDA Medical Device Recalls (openFDA)

Vygon MFG, Inc., dba/ Churchill Medical Systems, Inc.: Vygon IV Start Convenience Kits; Product Codes AMS-623T and AMS-626TGSLF. Kit used when administering IV during emergency.

Recall number
Z-0933-2018
Recalling firm
Vygon MFG, Inc., dba/ Churchill Medical Systems, Inc.
Product
Vygon IV Start Convenience Kits; Product Codes AMS-623T and AMS-626TGSLF. Kit used when administering IV during emergency.
Status
Terminated
Initiated
2017-10-03
Posted
not on file
Terminated
2020-01-06
Reason
The Iodine Prep Pads within the Vygon IV Start Convenience Kits, had stability samples that were not meeting iodine assay level requirements to support the 36 month expiration date. All kits containing the affected pads were recalled.
Root cause
Under Investigation by firm
Action
Vygon sent an Urgent Medical Device Recall letter dated October 3, 2017. Customers requested to check stock and quarantine affected devices and to complete Recall Acknowledgement and Inventory Return Form. Customers instructed to fax or email the form to Vygon including customer contact info. Vygon will contact the customer and issue a Return Merchandise Authorization Number. Distributors instructed to communicate recall to customers who received affected product and provide them copy of the recall notification. Returned products were received in Lansdale, PA warehouse location and placed into quarantine. For further questions, please call Customer Service at (800) 473-5414.
Quantity
21350
Distribution
US Distribution.
Lot, serial or model codes
AMS-623T-6 (1506075D, 1508057D, 1511030D, 1604024D, 1605080D, 1610017D, 1706097D, 1707055D) and AMS-626TGSLF (1601102D, 1603099D, 160581D, 1607019D, 1610057D, 1703068D, 1703069D, 1704065d, 1706076D, 1707031d, 1707056D)
510(k) numbers
["K882503"]
PMA numbers
not on file
Product code
LRS
Firm FEI
1000120555
Firm city
Dover
Firm state
NH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register