KCI USA, INC.: CelluTome Epidermal Harvesting System Harvester (5.0 CM) STERILE EO Rx Only Manufactured for: KCI USA, Inc. Part Number CT-H50 or CT-H25 The CelluTome Epidermal Harvesting System is intended to reproducibly cut a thin skin graft for autologous skin grafting.

Recall number
Z-0933-2015
Recalling firm
KCI USA, INC.
Product
CelluTome Epidermal Harvesting System Harvester (5.0 CM) STERILE EO Rx Only Manufactured for: KCI USA, Inc. Part Number CT-H50 or CT-H25 The CelluTome Epidermal Harvesting System is intended to reproducibly cut a thin skin graft for autologous skin grafting.
Status
Terminated
Initiated
2014-11-19
Posted
2015-01-07
Terminated
2015-09-18
Reason
KCI has received reports that, in a small number of cases, the CelluTome Harvester (a component of the CelluTome Epidermal Harvesting System), blades have the potential to drift during shipping and could prematurely protrude into openings of harvester top plate.
Root cause
Device Design
Action
KCI sent an Urgent - Voluntary Medical Device Field Notification letter to all affected consignees on December 8, 2014. The letter notified the customer of the issue, and provided instructions to inspect their harvester prior to use. If blade drift is identified, instructionswere provided to the customer to either return the device to KCI, or if they wish, how to reset the blade prior to use. The letter instructs the customer to respond to the recall notifcation by returning the product reconciliation form attached the letter. Follow-up notifcations will be send to non-responding consignees. Consignees with questions were instructed to contact their local KCI Representative or Customer Technical Service Center at 1-800-275-4524 and select option 3. For questions regarding this recall call 210-515-4108.
Quantity
1,139/5 Packs
Distribution
Worldwide Distribution - USA (Nationwide) and Internationally to Australia, Austria, Canada, France, Germany, Ireland, Italy, Isreal, Netherlands, South Africa, Spain, Sweden, Switerland, Turkey, UAE, and United Kingdom.
Lot, serial or model codes
CT-H25 lot numbers: 3249980000, 3249990000, 3250010000, 2537872, 2556866; CT-H50 lot numbers: 3249660000, 3249840000, 3250050000, 2498886, 2510874, 2537879, 2545871
510(k) numbers
not on file
PMA numbers
not on file
Product code
GFD
Firm FEI
3009897021
Firm city
San Antonio
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register