Skip to the content

FDA Medical Device Recalls (openFDA)

Fujifilm Medical Systems U.S.A., Inc.: Wako HDL-C/LDL-C Calibrator - Product Usage: designed to be used with Wako L-Type HDL-C or L-Type LDL-C for the determination of high density lipoprotein cholesterol (HDL-C) or low density lipoprotein cholesterol (LDL-C) in serum or plasma.

Recall number
Z-0932-2021
Recalling firm
Fujifilm Medical Systems U.S.A., Inc.
Product
Wako HDL-C/LDL-C Calibrator - Product Usage: designed to be used with Wako L-Type HDL-C or L-Type LDL-C for the determination of high density lipoprotein cholesterol (HDL-C) or low density lipoprotein cholesterol (LDL-C) in serum or plasma.
Status
Terminated
Initiated
2020-12-22
Posted
not on file
Terminated
2021-07-20
Reason
FUJIFILM has become aware that incorrect concentration values were printed on the bottle label for all product manufactured as part of Lot EL629. This may result incorrect results when used in testing.
Root cause
Under Investigation by firm
Action
On, 12/22/2020, FUJIFILM issued an Urgent Medical Device Removal notice to customers via email. Incorrect concentration values were printed on the bottle label for all product manufactured as part of Lot EL629. This may result in incorrect results when used in testing.
Quantity
10 boxes (4 units per box)
Distribution
US Nationwide distribution including in the states of MO, OR, MD, PA.
Lot, serial or model codes
Lot EL629, Model 990-28011
510(k) numbers
["K001005"]
PMA numbers
not on file
Product code
JIS
Firm FEI
1000513161
Firm city
Lexington
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register