Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Medical Systems (Cleveland) Inc: Diagnostic imaging systems (GEMINI LXL Model #882412, TruFlight Select Model #882438, GEMINI TF PET/CT16 Model #882470, GEMINI TF PET/CT 64 Model #882471, GEMINI TF Ready Model #882473 & GEMINI TF Big Bore Model #882476)

Recall number
Z-0932-2014
Recalling firm
Philips Medical Systems (Cleveland) Inc
Product
Diagnostic imaging systems (GEMINI LXL Model #882412, TruFlight Select Model #882438, GEMINI TF PET/CT16 Model #882470, GEMINI TF PET/CT 64 Model #882471, GEMINI TF Ready Model #882473 & GEMINI TF Big Bore Model #882476)
Status
Terminated
Initiated
2013-12-04
Posted
2014-02-04
Terminated
2016-02-02
Reason
The PET Reconstruction Server (PRS) database may randomly lock up after the low dose CT acquisition is complete and will not allow the PET acquisition to begin, resulting in an unexpected incomplete study.
Root cause
Software design
Action
On 12/4/2013, the firm sent Field Safety Noticies to their customers.
Quantity
386 units
Distribution
Product was shipped to the following states: AL, AR, AZ, CA, CO, DC, FL, GA, HI, IA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, NY , OH, OR, PA, TN, TX, VA, VT, WA & WI. Product was also shipped to the following countries: Australia, Belgium, Brazil, Bulgaria, Canada, China, Croatia, Cuba, Denmark, Egypt, France, Germany, Hong Kong, Hungary, India, Indonesia, Italy, Japan, Kuwait, Lebanon, Malta, Mexico, Netherlands, Pakistan, Philippines, Poland, Portugal, Republic of Korea, Russia, Russian Federation, Saudi Arabia, Singapore, Slovakia, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, Ukraine, United Kingdom, Venezuela & Vietnam.
Lot, serial or model codes
Model #882412; Serial #'s:14701, 14702, 14704, 14705, 14706, 14707, 14708, 14709, 14710 & 14711.  Model #882438; Serial #'s:17002, 17003, 17004, 17005, 17006, 17007, 17008, 17009, 17010, 17011, 17012, 17013, 17014, 17015, 17017, 17018, 17019, 17021, 17022, 17023, 17024, 17025, 17026 & 17028.  Model #882470; Serial #'s: 7003, 7004, 7005, 7006, 7008, 7009, 7010, 7011, 7014, 7015, 7017, 7019, 7020, 7021, 7022, 7024, 7025, 7026, 7031, 7034, 7035, 7036, 7039, 7043, 7045, 7049, 7050, 7052, 7053, 7054, 7058, 7060, 7062, 7064, 7066, 7068, 7069, 7074, 7075, 7077, 7080, 7081, 7084, 7085, 7086, 7087, 7089, 7093, 7094, 7095, 7096, 7097, 7098, 7099, 7108, 7112, 7114, 7115, 7118, 7120, 7121, 7122, 7123, 7124, 7125, 7128, 7129, 7131, 7135, 7137, 7138, 7139, 7140, 7142, 7143, 7145, 7147, 7148, 7156, 7159, 7161, 7163, 7166, 7168, 7173, 7174, 7176, 7177, 7181, 7183, 7184, 7185, 7186, 7187, 7193, 7194, 7195, 7196, 7197, 7198, 7199, 7200, 7207, 7208, 7210, 7211, 7215, 7216, 7223, 7224, 7226, 7228, 7230, 7233, 7235, 7239, 7240, 7242, 7501, 7503, 7507, 7508, 7511, 7514, 7517, 7518, 7520, 7521, 7523, 7527, 7528, 7529, 7530, 7531, 7532, 7533, 7536, 7538, 7540, 7545, 7546, 7554, 7559, 7564, 7566, 7568, 7570, 7573, 7581, 7583, 7584, 7586, 7590, 7591, 7104M, 7106M, 7170M &7582M.  Model #882471; Serial #'s: 7007, 7012, 7013, 7016, 7027, 7028, 7029, 7030, 7032, 7037, 7042, 7044, 7048, 7055, 7057, 7059, 7061, 7063, 7065, 7071, 7072, 7076, 7078, 7079, 7082, 7083, 7088, 7091, 7092, 7102, 7103, 7105, 7116, 7117, 7119, 7126, 7132, 7136, 7141, 7144, 7146, 7149, 7152, 7153, 7154, 7158, 7160, 7164, 7167, 7169, 7171, 7172, 7175, 7178, 7179, 7180, 7182, 7188, 7189, 7190, 7192, 7201, 7202, 7203, 7204, 7205, 7206, 7209, 7212, 7213, 7214, 7217, 7218, 7219, 7221, 7227, 7229, 7231, 7234, 7236, 7238, 7241, 7245, 7251, 7252, 7253, 7254, 7255, 7256, 7257, 7258, 7504, 7506, 7509, 7510, 7512, 7515, 7516, 7519, 7524, 7525, 7535, 7537, 7539, 7542, 7543, 7544, 7548, 7549, 7550, 7552, 7553, 7555, 7556, 7557, 7558, 7560, 7561, 7562, 7563, 7567, 7569, 7572, 7574, 7575, 7577, 7578, 7579, 7585, 7587, 7589 &7592.  Model #882473; Serial #'s: 7040, 7048, 7070, 7095, 7111, 7113, 7127, 7130, 7134, 7150, 7151, 7155, 7157, 7162, 7237 & 7249.  Model #:882476; Serial #'s: 9004, 9005, 9006, 9008, 9009, 9010, 9011, 9012, 9013, 9014, 9015, 9016, 9017, 9018, 9019, 9020, 9021, 9022, 9023, 9201, 9202, 9203, 9204, 9205, 9206, 9207, 9208, 9209, 9210, 9211, 9213, 9214, 9215, 9216, 9217, 9218, 9219, 9220, 9221, 9222, 9223, 9224, 9225, 9226, 9228, 9229 & 9231.
510(k) numbers
["K052640","K081135"]
PMA numbers
not on file
Product code
KPS
Firm FEI
1525965
Firm city
Cleveland
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register