Skip to the content

FDA Medical Device Recalls (openFDA)

Alphatec Spine, Inc.: Solanas Posterior Stabilization System Part Number: 63920. Solanas Posterior Stabilization System facilitates the surgical correction of spinal deformities by providing temporary internal fixation and stabilization during bone graft healing and/or fusion mass development. The Solanas Posterior Stabilization System is a spinal fixation system intended to improve stability of the cervical and thoracolumbar area of the spine (C1-T3).

Recall number
Z-0932-2012
Recalling firm
Alphatec Spine, Inc.
Product
Solanas Posterior Stabilization System Part Number: 63920. Solanas Posterior Stabilization System facilitates the surgical correction of spinal deformities by providing temporary internal fixation and stabilization during bone graft healing and/or fusion mass development. The Solanas Posterior Stabilization System is a spinal fixation system intended to improve stability of the cervical and thoracolumbar area of the spine (C1-T3).
Status
Terminated
Initiated
2010-11-01
Posted
2012-01-31
Terminated
2012-01-31
Reason
These lots were not manufactured correctly in that the instrument does not have the durability required to repeatedly perform over its full service life and, therefore, may experience breakage in the tips.
Root cause
Device Design
Action
Alphatec Spine contacted all affected customers via telephone. The communication included discussion of the product, problem, and actions to be taken by the customers. Customers were instructed to remove the affected product from inventory and to return the product to the firm. Arrangements were made for replacement product to be sent to the customers. Contact the firm at 1-800-922-1356 for questions regarding this recall.
Quantity
27 units
Distribution
Nationwide Distribution-including the states of CA, CT, FL, GA, NC, NE, NJ, TX, and UT.
Lot, serial or model codes
Lot Numbers: 608227B, 609867B, 609870B, 609882B, 609897B, 609971B
510(k) numbers
["K052201","K071380"]
PMA numbers
not on file
Product code
KWP
Firm FEI
2027467
Firm city
Carlsbad
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register