Skip to the content

FDA Medical Device Recalls (openFDA)

Ion Beam Applications S.A.: Proteus 235, Pencil Beam Scanning version PTS-6.4.11.X The Proton Therapy System - Proteus 235 (brand names: Proteus Plus and Proteus ONE) is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation. The PTS may include a fixed small beam treatment room dedicated to the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation localized to the head and neck.

Recall number
Z-0931-2018
Recalling firm
Ion Beam Applications S.A.
Product
Proteus 235, Pencil Beam Scanning version PTS-6.4.11.X The Proton Therapy System - Proteus 235 (brand names: Proteus Plus and Proteus ONE) is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation. The PTS may include a fixed small beam treatment room dedicated to the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation localized to the head and neck.
Status
Terminated
Initiated
2017-10-19
Posted
not on file
Terminated
2019-04-02
Reason
Range for the same energy could be different depending if the layer is the first layer of the irradiation plan or not.
Root cause
Device Design
Action
IBA sent an Urgent Field Safety Notice dated October 19, 2017 , issued to customers stating the following: IBA recommends performing a check to detect the consequence of this potential power supply regulation issue on a regular basis (IBA recommends Monthly), see Annex 1 for details. IBA is developing a corrective action in order to reduce the hysteresis effect and detect magnet current overshoot. This solution will be deployed to all systems using JEMA analog power supply. This solution will be validated and deployed by April 2018. For further questions, please call (215) 972-7985
Quantity
3
Distribution
US Distribution to the state of :FL
Lot, serial or model codes
PAT.003 (KR), PAT.006 (US)
510(k) numbers
["K082416"]
PMA numbers
not on file
Product code
LHN
Firm FEI
3000256071
Firm city
Louvain La Neuve
Firm state
not on file
Firm country
Belgium

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register