Skip to the content

FDA Medical Device Recalls (openFDA)

Elekta, Inc.: Elekta Synergy XVI. Radiation Therapy Digital Imager. Product XVI R3.5.1, R4.2.1, and R4.5.1 Used as part of radiation therapy treatment process.

Recall number
Z-0931-2013
Recalling firm
Elekta, Inc.
Product
Elekta Synergy XVI. Radiation Therapy Digital Imager. Product XVI R3.5.1, R4.2.1, and R4.5.1 Used as part of radiation therapy treatment process.
Status
Terminated
Initiated
2012-12-19
Posted
2013-03-10
Terminated
2021-07-14
Reason
The DICOM tags do not contain sufficient information for XVI to identify the different volumes that have the same name.
Root cause
Software design
Action
Elekta sent an "IMPORTANT FIELD SAFETY NOTICE" dated December 2012 to all affected customers. The letter identifies the product, problem, and the actions to be taken by the customers. The IFSN informs the user in the TPS, to give the different volumes different structure names. After importing the patient data set from the TPS, make sure that the structure contours are correct. Users are requested to return the signed acknowledgement. Contact your local Elekta office for questions regarding this notice.
Quantity
164
Distribution
Nationwide Distribution-DC, Puerto Rico and the states of AL, AK, AZ, CA, CO, CT, FL, GA, IL, IA, KY, MD, MA, MI, MN, MS, MT, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, SC, TN, TX, VT, VA, WA, WV, and WI.
Lot, serial or model codes
151055, 151700, 151809, 151695, 151479, 151696, 151867, 151704, 151765, 151860, 151024, 151521, 151803, 152217, 151774, 152220, 151626, 151842, 151847, 151446, 151801, 151953, 151664, 151751, 151352, 151298, 151728, 151504, 151981, 151705, 151977, 152123, 152126, 151955, 105982, 151160, 151634, 151954, 151416, 151074, 151587, 151729, 151694, 151965, 151418, 151674, 151259, 151698, 151699, 152043, 151636, 151811, 151568, 151258, 151584, 151423, 151623, 151627, 151507, 151640, 151933, 151340, 151722, 151810, 151051, 151411, 151360, 151595, 151433, 151611, 152008, 151151, 151714, 151653, 151675, 151538, 152031, 151523, 151519, 151432, 151783, 151410, 151371, 151084, 151387, 151978, 151530, 151328, 151689, 151567, 151598, 151760, 151269, 151557, 151558, 151789, 151398, 151632, 152214, 151301, 151441, 151256, 151391, 151038, 151585, 151327, 151326, 151250, 151130, 151828, 151245, 151672, 105803, 151073, 151956, 152232, 151319, 151959, 151960, 151958, 151435, 151779, 151969, 151848, 151875, 151881, 151882, 151736, 152023, 152080, 152079, 152064, 151734, 151763, 152154, 151110, 151363, 151216, 151173, 151678, 151467, 151961, 151178, 152091, 152051, 152115, 152116, 151708, 152176, 152192, 152098, 152200, 152174, 152246, 152364, 152673, 152425, 151156, 151764, 151886, 151076, 151369, 152175, 105941
510(k) numbers
["K051932"]
PMA numbers
not on file
Product code
IYE
Firm FEI
1037831
Firm city
Atlanta
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register