Elekta, Inc.: Elekta Synergy XVI. Radiation Therapy Digital Imager. Product XVI R3.5.1, R4.2.1, and R4.5.1 Used as part of radiation therapy treatment process.
- Recall number
- Z-0931-2013
- Recalling firm
- Elekta, Inc.
- Product
- Elekta Synergy XVI. Radiation Therapy Digital Imager. Product XVI R3.5.1, R4.2.1, and R4.5.1 Used as part of radiation therapy treatment process.
- Status
- Terminated
- Initiated
- 2012-12-19
- Posted
- 2013-03-10
- Terminated
- 2021-07-14
- Reason
- The DICOM tags do not contain sufficient information for XVI to identify the different volumes that have the same name.
- Root cause
- Software design
- Action
- Elekta sent an "IMPORTANT FIELD SAFETY NOTICE" dated December 2012 to all affected customers. The letter identifies the product, problem, and the actions to be taken by the customers. The IFSN informs the user in the TPS, to give the different volumes different structure names. After importing the patient data set from the TPS, make sure that the structure contours are correct. Users are requested to return the signed acknowledgement. Contact your local Elekta office for questions regarding this notice.
- Quantity
- 164
- Distribution
- Nationwide Distribution-DC, Puerto Rico and the states of AL, AK, AZ, CA, CO, CT, FL, GA, IL, IA, KY, MD, MA, MI, MN, MS, MT, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, SC, TN, TX, VT, VA, WA, WV, and WI.
- Lot, serial or model codes
- 151055, 151700, 151809, 151695, 151479, 151696, 151867, 151704, 151765, 151860, 151024, 151521, 151803, 152217, 151774, 152220, 151626, 151842, 151847, 151446, 151801, 151953, 151664, 151751, 151352, 151298, 151728, 151504, 151981, 151705, 151977, 152123, 152126, 151955, 105982, 151160, 151634, 151954, 151416, 151074, 151587, 151729, 151694, 151965, 151418, 151674, 151259, 151698, 151699, 152043, 151636, 151811, 151568, 151258, 151584, 151423, 151623, 151627, 151507, 151640, 151933, 151340, 151722, 151810, 151051, 151411, 151360, 151595, 151433, 151611, 152008, 151151, 151714, 151653, 151675, 151538, 152031, 151523, 151519, 151432, 151783, 151410, 151371, 151084, 151387, 151978, 151530, 151328, 151689, 151567, 151598, 151760, 151269, 151557, 151558, 151789, 151398, 151632, 152214, 151301, 151441, 151256, 151391, 151038, 151585, 151327, 151326, 151250, 151130, 151828, 151245, 151672, 105803, 151073, 151956, 152232, 151319, 151959, 151960, 151958, 151435, 151779, 151969, 151848, 151875, 151881, 151882, 151736, 152023, 152080, 152079, 152064, 151734, 151763, 152154, 151110, 151363, 151216, 151173, 151678, 151467, 151961, 151178, 152091, 152051, 152115, 152116, 151708, 152176, 152192, 152098, 152200, 152174, 152246, 152364, 152673, 152425, 151156, 151764, 151886, 151076, 151369, 152175, 105941
- 510(k) numbers
- ["K051932"]
- PMA numbers
- not on file
- Product code
- IYE
- Firm FEI
- 1037831
- Firm city
- Atlanta
- Firm state
- GA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28