Siemens Healthcare Diagnostics, Inc: ADVIA Centaur Systems Phenytoin Calibrator N ( PHTN) 6-pack SMN 10314977 For in vitro diagnostic use in the quantitative determination of phenytoin in serum or plasma using the ADVIA Centaur and ADVIA Centaur XP systems

Recall number
Z-0930-2014
Recalling firm
Siemens Healthcare Diagnostics, Inc
Product
ADVIA Centaur Systems Phenytoin Calibrator N ( PHTN) 6-pack SMN 10314977 For in vitro diagnostic use in the quantitative determination of phenytoin in serum or plasma using the ADVIA Centaur and ADVIA Centaur XP systems
Status
Terminated
Initiated
2013-12-18
Posted
2014-02-04
Terminated
2016-05-18
Reason
ADVIA Centaur Systems Phenytoin Calibrator N, lots CN58 and higher, change in recovery observed with the ADVIA Centaur Systems Phenytoin assay compared to recoveries using previous Calibrator N lots.
Root cause
Under Investigation by firm
Action
Siemens Healthcare Diagnostics Customer Notification notice was sent to affected Siemens Healthcare Diagnostic customers in the United States on December 18, 2013.. The notifications inform customers that Siemens Healthcare Diagnostics received customer feedback regarding the performance of the ADVIA Centara Systems Phenytoin assay, indicating that the assay was recovering higher relative to the performance of other assays in proficiency studies. Siemens internal investigation confirmed the positive bias for Calibrator N lot CN 57 in the ADVIA Centaur Systems Phenytoin assay. This lot has expired and is no longer in use. Expected performance has been restored starting with Calibrator N lot CN58. Siemens is notifying customers that have received Calibrator N lots CN58 and higher that there will be a change in recovery observed with the ADVIA Centaur Systems Phenytoin assay compared to recoveries using previous Calibrator N lots. Siemens is providing data to inform customers of the expected difference between Calibrator lots CN58 and higher versus prior Calibrator N lots. Customers were instructed to retain the letter with their laboratory records, and forward to those who may have received the affected product. Customers with questions were instructed to contact their Siemens Customers Care Center or their local Siemens technical support representative. For questions regarding this recall call 508-668-5000.
Quantity
1294 (144 units US)
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Argentina, Australia,Brazil,Canada, Columbia, China, Ecuador, Egypt, India, Japan, Korea, Mexico, Saudi Arabia,Singapore, Thailand, and Uruguay.
Lot, serial or model codes
Lot # Expiry Date  18998A58 06/09/2014; 21854A58 06/09/2014; 25487A59 09/29/2014.
510(k) numbers
["K011393"]
PMA numbers
not on file
Product code
DIP
Firm FEI
1219913
Firm city
East Walpole
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register