FDA Medical Device Recalls (openFDA)
Terumo Cardiovascular Systems Corp: Terumo Advanced Perfusion System 1; 100V -120V, 15 A (circuit breaker), 50/60 Hz (20A power source required) base; Model 801763.
- Recall number
- Z-0929-2007
- Recalling firm
- Terumo Cardiovascular Systems Corp
- Product
- Terumo Advanced Perfusion System 1; 100V -120V, 15 A (circuit breaker), 50/60 Hz (20A power source required) base; Model 801763.
- Status
- Terminated
- Initiated
- 2005-02-01
- Posted
- 2007-06-30
- Terminated
- 2009-07-07
- Reason
- System 1 power supply may electrically short to the main console, disrupting normal system operation by causing unexpected re-setting or shutdown of pumps and modules, false alarms and/or overheating and failure of the central control monitor.
- Root cause
- Other
- Action
- The firm corrected the complaint units at the time of the complaints in 2005. The firm issued a Device Correction Letter dated 6/6/07 notifying all users of the problem and advising them that their system will be corrected when replacement components become available.
- Quantity
- 416 for both models
- Distribution
- Worldwide: USA, Australia, Belgium, Canada, Dominican Republic, Egypt, Germany, Hong Kong, India, Indonesia, Japan, Malaysia, Mexico, Pakistan and Singapore.
- Lot, serial or model codes
- Serial numbers 0011 through 0039, 0041 through 0308, 0310 through 0313, 0315, 0316, 0319 through 0321 and 0324 may be affected by this issue. Serial numbers 0238 and 0282 have been corrected to date.
- 510(k) numbers
- ["K022947"]
- PMA numbers
- not on file
- Product code
- DTQ
- Firm FEI
- 1828100
- Firm city
- Ann Arbor
- Firm state
- MI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28