Skip to the content

FDA Medical Device Recalls (openFDA)

Datascope Corp: Anestar and Anestar Plus Anesthesia System.

Recall number
Z-0929-06
Recalling firm
Datascope Corp
Product
Anestar and Anestar Plus Anesthesia System.
Status
Terminated
Initiated
2006-03-17
Posted
2006-06-02
Terminated
2006-08-01
Reason
The Anestar and Anestar Plus Anesthesia System may produce periods of high Positive End-Expiratory Pressure (PEEP), in both manual and mechanical ventilation mode due to an issue with the Applied Pressure Limit (APL) valve.
Root cause
Other
Action
Field correction communications were sent 3/17/2006 to all accounts via certified mail, return receipt requested, signature required.
Quantity
337 UNITS
Distribution
The units have been distributed to distributors and end-users such as hospitals and clinics nationwide. There are 3 foreign accounts in Australia, El Salvador, and Venezuela.
Lot, serial or model codes
Serial Numbers starting with 4600: A51083; C51104; C51116; C51117; C51120; C51121; C51125; E41025; G30143; L41062; L41063; L41064; L41065; L41071; L41075; 010001; 010002; 010008; 010010; A20024; A51076; A51077; A51078; A51081; A51084; B20027; B30076; B30077; B30078; B30081; B30082; B30083; B41005; B51087; B51088; B51089; B51092; B51096; C30087; C30092; C30093; C30094; C30095; C30096; C30097; C30098; C30099; C30102; C41014; C41015; C41017; C41020; C51114; C51118; D20035; D20036; D20037; D20038; D20042; D20043; D20044; D30100; D30101; D30103; D30104; D30105; D30106; D30107; D30108; D30109; E30112; E30113; E30115; E30116; E30117; E30118; E41029; F30120; F30128; F30129; F30131; F30132; F30133; F30134; F30135; G30136; G30137; G30138; G30139; G30140; G30142; H20045; H20047; H20051; H20052; I10010; I41031; I41035; I41036; I41038; I41041; I41042; I41043; I41044; I41045; I41046; I41047; I41048; I41049; I41050; J20055; J20056; J20057; J20059; J20060; J20061; J20063; J20064; K41057; L41066; L41073; A20018; A20023; A30069; A30070; B20026; B20028; B20030; B30084; B30086; C41031; C41018; D20014; I41034; I41037; K41060; L30162; L41068; C51101; C51103; C51111; C51113; D51131; F30122US; F30127; H30149; I41033; I41039; I41040; 010007; E30114; H30144; H30145; H30146; H30147; H30148; H30150; H30151; H30152; H30153; H30155; A30066; A30067; A30071; A51080; A51082; A51085; B41002; B51086; B51090; B51097; C41011; C41061; C41091; C51105; C51108; C51115; C51123; D51127; D51129; D51132; D51135; E41022; E41023; E41024; E41026; F30130; H20048; I41032; K41051; K41052; K41053; K41054; K41059; L41069; A51079; B41004; B41009; A30072; A30073; A30074; A30075; D20039; D20040; E30119; H30154; 010012; A20020; A20034; K41055; K41056; K41058; L30156; L30157; L30159; L30160; L30161; L41067; L41070; L41074; A30068; B51091; B51093; B51094; B51095; C30088; C30089; C30090; C30091; D10005; E41021; E41027; E41028; E41030; L30163; L30167; B51098; C51112; B20032; D51126; D51130; D51133; D51134; L30164; L30166; F30121; F30123; A20015; A20016; A20022; B10001; B20029; B20031; B20033; B30079; B30080; B41001; B41003; B41006; B41007; B41008; B41010; C51102; C51107; C51109; C51110; F30124; F30125; F30126; G30141; H20046; H20049; H20050; H20053; H20054; I10008; I10013; J20058; J20062; L30158; L40161; A20021; L41072.  Serial Numbers starting with AF0: 1040-L5; 1045-A6; 1043-L5; 1029-K5; 1037-L5; 1005-I5; 1032-K5; 1036-L5; 1035-L5; 1034-L5; 1033-L5; 1032-L5; 1031-K5; 1030-K5; 1028-K5; 1027-K5; 1026-K5; 1025-K5; 1024-K5; 1023-K5; 1022-K5; 1021-K5; 1020-K5; 1019-K5; 1018-K5; 1017-K5; 1016-K5; 1015-K5; 1014-J5; 1013-J5; 1012-J5; 1011-J5; 1009-I5; 1008-I5; 1007-I5; 1004-I5; 1003-I5; 1001-I5; 1055-A6; 1051-A6; 1048-A6; 1046-A6; 1042-L5; 1010-I5; 1044-A6; 1041-L5; 1038-L5; 1006-I5; 1053-A6; 1052-A6; 1049-A6; 1047-A6; 1039-L5.
510(k) numbers
["K001988"]
PMA numbers
not on file
Product code
BSZ
Firm FEI
2221819
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register