Skip to the content

FDA Medical Device Recalls (openFDA)

Biomerieux Inc: BIOMERIEUX VIDAS 3, Compact immunoanalyzer with full traceability and automation

Recall number
Z-0928-2020
Recalling firm
Biomerieux Inc
Product
BIOMERIEUX VIDAS 3, Compact immunoanalyzer with full traceability and automation
Status
Terminated
Initiated
2019-10-22
Posted
not on file
Terminated
2024-08-06
Reason
Following Customers complaints, investigations have been initiated on potentially false results obtained on VIDAS 3 with an expired calibration. Indeed, the calibrations of assays were valid in the calibration menu whereas in fact the calibrations were expired and no alarm displayed to warn the users about the expiration date of the calibrations. Software computes results of analysis for assays with expired calibrations. The anomaly is due to an incorrect update of the calibration status by the software.
Root cause
Software design
Action
The firm initiated their field action by an email to their subsidiaries on 10/22/2019. The subsidiaries then notified their customers by 2-day express via Fed Ex. In the US customer letter, customers are instructed to check their software version. If the VIDAS 3 software version is between 1.2.0 and 1.3.1, customers are instructed to check the archiving mode. If the archiving mode is Automatic When Results are Reported then customers are instructed how to configure the workaround, and to contact bioMerieux if the customer has the incorrect software configuration. bioMerieux then instructs impacted customers to review and interpret results in the context of the overall clinical procedure. Once this is completed to complete the acknowledgment form in Attachment A and return it the local bioMerieux representative and confirm receipt of this notice.
Quantity
738 units
Distribution
US
Lot, serial or model codes
software versions 1.2 and higher
510(k) numbers
["K141133"]
PMA numbers
not on file
Product code
JJE
Firm FEI
1950204
Firm city
Hazelwood
Firm state
MO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06