Plus Orthopedics USA: Model Number: GNRADIOLUCENTGUIDE (Gliding Nail Radiolucent Guide), instrument set that contains one unit of each of the following instrument: 1130020-A: GNp Carbon Targeting Handle 1130021-A: GN Adapter Screw for Handle SW 10 1130022-A: GN Targeting Guide Carbon 125 1130023-A: GN Targeting Guide Carbon 135 1130060-A: GNp Drill Guide Locking No. 3 1133158-A: GN Centering Sleeve New Guide Description: Surgical instruments comprising the Gliding Nail Radiolucent Guide, supplied unsterile for use in reusable sterilization trays.

Recall number
Z-0928-06
Recalling firm
Plus Orthopedics USA
Product
Model Number: GNRADIOLUCENTGUIDE (Gliding Nail Radiolucent Guide), instrument set that contains one unit of each of the following instrument: 1130020-A: GNp Carbon Targeting Handle 1130021-A: GN Adapter Screw for Handle SW 10 1130022-A: GN Targeting Guide Carbon 125 1130023-A: GN Targeting Guide Carbon 135 1130060-A: GNp Drill Guide Locking No. 3 1133158-A: GN Centering Sleeve New Guide Description: Surgical instruments comprising the Gliding Nail Radiolucent Guide, supplied unsterile for use in reusable sterilization trays.
Status
Terminated
Initiated
2006-02-22
Posted
2006-06-02
Terminated
2007-12-13
Reason
PLUS USA has been informed by PLUS AG that the accuracy of the instrument set may be affected in some cases. It is possible that in rare cases where patients have sclerotic (excessively hardened or firm) bone, that components may deform under application of force as the surgeon attempts to penetrate this bone with the instruments.
Root cause
Other
Action
There were a total of five (5) radiolucent guide instrument sets affected by the recall (imported to the U.S. between 1/23/2006 and 1/30/2006). The firm telephoned agents beginning February 22, 2006, requesting return of the devices. Returned Material Authorizations have been issued (March, 2006) for consigned inventory in the field with instructions given to remove and return affected units immediately. All returns to PLUS Orthopedics USA have been appropriately quarantined and segregated to prevent unintended use or shipment.
Quantity
5
Distribution
There were a total of five (5) radiolucent guide instrument sets affected by the recall (imported to the U.S. between 1/23/2006 and 1/30/2006). The 5 sets were distributed in the following states: 1) California, 2) Idaho, and 3) New York.
Lot, serial or model codes
Revision A Serial Number: None
510(k) numbers
not on file
PMA numbers
not on file
Product code
KTT
Firm FEI
3001237832
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register