Becton Dickinson & Company: BD 25G Spinal Needles. (Spinal Anesthesia Needle)

Recall number
Z-0928-05
Recalling firm
Becton Dickinson & Company
Product
BD 25G Spinal Needles. (Spinal Anesthesia Needle)
Status
Terminated
Initiated
2005-03-16
Posted
2005-06-25
Terminated
2005-11-07
Reason
A significant level of 25G spinal needle hub leakage complaints was observed. Subsequent investigation found the leakage to be attributable to improper formation of the spinal needle hub, specifically an undersized taper and the prescence of a slight sink in the taper surface.
Root cause
Other
Action
The firm sent out notification letters on 3/16/2005.
Quantity
801580 units
Distribution
The firm uses distributors and ships the product directly to over 4000 hospitals and clinics nationwide.
Lot, serial or model codes
Reorder Number, Lot Number: 400440, 4275101; 400440, 4282078; 400440, 4317744; 400479, 3196624; 400479, 4161815; 400479, 4254414; 400479, 4324795; 400496, 4173091; 400590, 3191978; 400590, 4121350; 400621, 4146482; 401586, 3182399; 405078, 3196618; 405078, 4169804; 405078, 4194109; 405078, 4268754; 405078, 4282079; 405078, 4324791; 405078, 4349021; 405138, 4173087; 405138, 4181398; 405140, 3224096; 405140, 4110295; 405140, 4237191; 405140, 4295872, 405140, 4314481; 405140, 4336365; 405170, 4201356; 405170, 4201357 405180, 4096782; 405180, 4145523; 405180, 4148675; 405234, 3174061; 405234, 4062807; 405234, 4105111; 405234, 4163586; 405234, 4265265; 405234, 4282715; 405234, 4282716; 405234, 4314480; 405234, 4358850; 409442, 3174062; 409442, 3226365; 409442, 4110275; 409442, 4121348; 409442, 4230810; 409442, 4230813; 409442, 4257418; 409442, 4300308; 409442, 4320423; 409442, 4329215; 409442, 4349941.
510(k) numbers
not on file
PMA numbers
not on file
Product code
BSP
Firm FEI
2243072
Firm city
Franklin Lakes
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register