Skip to the content

FDA Medical Device Recalls (openFDA)

DeRoyal Industries Inc: DeRoyal ORGAN RECOVERY PACK PGYBK, a) REF 89-9004.09 and b) 89-9004.10

Recall number
Z-0927-2023
Recalling firm
DeRoyal Industries Inc
Product
DeRoyal ORGAN RECOVERY PACK PGYBK, a) REF 89-9004.09 and b) 89-9004.10
Status
Terminated
Initiated
2022-12-12
Posted
2023-01-09
Terminated
2026-05-29
Reason
DeRoyal is recalling certain lots of surgical procedure pack products that contain a Medtronic Non-Absorbable Suture. The recall is due to Medtronic determining a manufacturing error in the packaging of the non-absorbable sutures that may cause a gap or wrinkle in the breather seal resulting in a sterile barrier breach.
Root cause
Packaging change control
Action
DeRoyal sent an urgent recall notice to its consignees on 12/12/2022 by email and they followed with a letter sent overnight to the end user. The notice explained the problem and the hazard, and requested the following actions be taken: "1) Using the attached Affected Products Listing, identify affected surgical pack numbers in your inventory and place in quarantine to prevent further use. Please return the packs that you have in inventory to us. You will be given credit or replacement for the returned product. 2) Complete the NOTICE OF RETURN FORM, indicating all affected products found in your inventory. Please return the completed form to DeRoyal via fax to 865-362-3716 or recalls@deroyal.com no later than January 15, 2023. 3) If you have questions regarding credit for the recall, please contact Jennifer Marsh at jmarsh@deroyal.com or by phone 865-362-4203 or email us at recalls@deroyal.com. 4) PLEASE COMPLETE AND RETURN THIS FORM EVEN IF YOU NO LONGER HAVE ANY INVENTORY. 5) If you have further distributed these products, it is required by federal law that your customers be notified of this recall. You may either do this directly by forwarding our recall notification and notify us that you have done so, or you may provide us with your customer listing and we will contact them."
Quantity
157 packs
Distribution
US Nationwide distribution in the states of IL, MO.
Lot, serial or model codes
a) REF 89-9004.09, GTIN 00749756903936, Lot Numbers: 57391129 exp 5/1/2023, 57755421 exp 5/1/2023, 57900931 exp 5/1/2023 b) REF 89-9004.10, GTIN 00749756903931, Lot Numbers: 57913767 exp 5/1/2023
510(k) numbers
["K842648"]
PMA numbers
not on file
Product code
LRO
Firm FEI
1043214
Firm city
Powell
Firm state
TN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register