Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: AXIOM Artis and AXIOM Artis zee modular angiographic systems Product Usage: Angiographic x-ray systems

Recall number
Z-0927-2012
Recalling firm
Siemens Medical Solutions USA, Inc
Product
AXIOM Artis and AXIOM Artis zee modular angiographic systems Product Usage: Angiographic x-ray systems
Status
Terminated
Initiated
2012-01-16
Posted
2012-01-27
Terminated
2014-01-31
Reason
There is a potential misalignment in the connection of the remote foot switch and the stationary unit. This potential malfunction may occur if either the wireless foot switch or the stationary unit is equipped with improper firmware installed in the bluetooth interface performing the communication between foot switch and system. This is not a general issue but sporadic as a vendor unsystematical
Root cause
Software change control
Action
Siemens sent an Urgent Field Safety Notice letter on January 16, 2016, to all affected customers The letter identified the product, problem, potential risk, and actions to be taken. Customers were instructed to immediately pass this information to all those who need to be aware and instruct personnel accordingly. Siemens ask that if the device/equipment is no longer in the customers possession to forward this safety notice to the new owner of the device/equipment.
Quantity
45
Distribution
Nationwide Distribution (USA) including the states of: CA, CT, FL, GA, HI, IL, IN, KY, ME, MD, MI, MS, MO, NH, NJ, NY, NC, OH, OR, TN, TX, UT, VA, WI and Puerto Rico.
Lot, serial or model codes
Model numbers 5904433, 5904441, 5904649, 5904656, 5904466, 7007755, 7728350, 7412807, 7727717, 7728392, 7413078, 7555373, 7555365, 7555357, 7008605, 5917054, 10094135, 10094137, 10094139, 10094141, 10094142, 10094143, and 10280959.
510(k) numbers
["K052202"]
PMA numbers
not on file
Product code
JAA
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register