Skip to the content

FDA Medical Device Recalls (openFDA)

Exactech, Inc.: Product is Optetrak Cemented Finned Tibial Tray Sz. 4F/5T. Knee prosthesis. Product is labeled in part: ''OPTETRAK FINNED TIBIAL TRAY A/P53 mm M/L 83mm***STERILE, SINGLE USE ONLY FOR CEMENTED USE ONLY***EXACTECH Gainesville, FL 32653***. ''

Recall number
Z-0927-06
Recalling firm
Exactech, Inc.
Product
Product is Optetrak Cemented Finned Tibial Tray Sz. 4F/5T. Knee prosthesis. Product is labeled in part: ''OPTETRAK FINNED TIBIAL TRAY A/P53 mm M/L 83mm***STERILE, SINGLE USE ONLY FOR CEMENTED USE ONLY***EXACTECH Gainesville, FL 32653***. ''
Status
Terminated
Initiated
2005-12-19
Posted
2006-06-02
Terminated
2012-06-29
Reason
An implant within a lot of finned tibial trays manufactured at Exactech was not made to specification. The locking cavity in the tibial tray was not fully machined thus preventing the tibial insert from locking in place.
Root cause
Other
Action
Each consignee contacted by mail 12/19/2005, and asked to return the product to Exactech. The returned product will be quarantined and evaluated.
Quantity
14 implants from manufactured lot
Distribution
Distributed domestically to Alabama and Louisiana, and internationally to Spain, Germany and Greece.
Lot, serial or model codes
Catalogue number 200-04-45; Serial number range 0730621-07306351; Lot number 452500.
510(k) numbers
["K932776"]
PMA numbers
not on file
Product code
JWH
Firm FEI
1038671
Firm city
Gainesville
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register