Skip to the content

FDA Medical Device Recalls (openFDA)

ArjoHuntleigh Inc.: The Rotoprone therapy System is an advanced patient care system for the treatment and prevention of pulmonary complications. The Rotoprone Therapy System provides up to 62 degrees in both prone and supine positions as well as Trendelenburg/Reverse Trendelenburg positioning. Product Usage: Use of the Rotoprone Therapy System is indicated for patients with severe acute respiratory distress syndrome (ARDS). These are critical patients who are at higher risk of cardiac arrest and mortality. Placing a patient in the prone position may reopen collapsed airways and thus improve pulmonary gas exchange and respiratory mechanics.

Recall number
Z-0926-2014
Recalling firm
ArjoHuntleigh Inc.
Product
The Rotoprone therapy System is an advanced patient care system for the treatment and prevention of pulmonary complications. The Rotoprone Therapy System provides up to 62 degrees in both prone and supine positions as well as Trendelenburg/Reverse Trendelenburg positioning. Product Usage: Use of the Rotoprone Therapy System is indicated for patients with severe acute respiratory distress syndrome (ARDS). These are critical patients who are at higher risk of cardiac arrest and mortality. Placing a patient in the prone position may reopen collapsed airways and thus improve pulmonary gas exchange and respiratory mechanics.
Status
Terminated
Initiated
2013-09-11
Posted
2014-02-04
Terminated
2015-01-08
Reason
The lock pin can become stuck in the "in" position during the manual rotation, resulting in the inability to prone a patient
Root cause
Device Design
Action
ArjoHuntleigh Inc. sent a Technical Bulletin package to all affected service centers and an Urgent Field Safety Notice package to all affected customers. For rental units, a training plan was established for employees. For sold units, the notice contain a revised user manual and label explaining what to do if the lock pin becomes stuck in the "in" position. Customers that purchased the affected product were required to complete and return a Customer Response form to ArjoHuntleigh to confirm receipt and understanding of the field safety notice, and that the required actions were completed. For questions call 630-785-4885.
Quantity
231 units
Distribution
US Nationwide and Middle East
Lot, serial or model codes
Model #209500
510(k) numbers
not on file
PMA numbers
not on file
Product code
IKZ
Firm FEI
3010048749
Firm city
San Antonio
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register