FDA Medical Device Recalls (openFDA)
Stryker Instruments Div. of Stryker Corporation: Stryker PainPump 2 Blockaid, 400 mL PainPump for Continuous Nerve Block, sterile, 5 per box. Stryker, Kalamazoo, MI; REF 575-100, Product Number 0575-100-000. The device is used to deliver controlled amounts of medication and narcotics directly to the intraoperative site for pain management, the pump infuses the medication at an hourly flow rate and provides the option for patient controlled bolus doses. Medications and narcotics are infused through intramuscular, subcutaneous and epidural routs. It is also intended for controlled delivery of local anesthetics and narcotics in close proximity to nerves for postoperative regional anesthesia and pain management, in which case routes of administration may be intraoperative, perineural or percutaneous.
- Recall number
- Z-0926-2009
- Recalling firm
- Stryker Instruments Div. of Stryker Corporation
- Product
- Stryker PainPump 2 Blockaid, 400 mL PainPump for Continuous Nerve Block, sterile, 5 per box. Stryker, Kalamazoo, MI; REF 575-100, Product Number 0575-100-000. The device is used to deliver controlled amounts of medication and narcotics directly to the intraoperative site for pain management, the pump infuses the medication at an hourly flow rate and provides the option for patient controlled bolus doses. Medications and narcotics are infused through intramuscular, subcutaneous and epidural routs. It is also intended for controlled delivery of local anesthetics and narcotics in close proximity to nerves for postoperative regional anesthesia and pain management, in which case routes of administration may be intraoperative, perineural or percutaneous.
- Status
- Terminated
- Initiated
- 2008-11-12
- Posted
- 2009-01-27
- Terminated
- 2009-11-30
- Reason
- The pump may display an E-4 error message, which causes the pump to stop dispensing the pain medication.
- Root cause
- Process design
- Action
- Consignees were notified via letter dated 11/12/08 and instructed to locate and destroy the affected lots. Customers were also instructed to complete and return the enclosed Business Reply form via fax (336-499-8603) to Stryker Instruments Regulatory Department. For questions regarding the recalled product, please contact Stryker Instruments at 1-800-800-4236, extension 3808 or extension 4056.
- Quantity
- 2,009 boxes
- Distribution
- Worldwide Distribution including USA and country of Canada.
- Lot, serial or model codes
- Lot Numbers: 08219012, 08220012, 08221012, 08222012, 08225012, 08226012, 08229012, 08231022, 08233012, 08234012, 08235012, 08238022, 08239012, 08241012, 08242012, 08243012, 08246022, 08247012, 08249032, 08250012, 08252022, 08252032, 08253012, 08254012, 08256032, 08256042, 08257012 and 08259022.
- 510(k) numbers
- ["K043466"]
- PMA numbers
- not on file
- Product code
- FRN
- Firm FEI
- 1811755
- Firm city
- Portage
- Firm state
- MI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28