Cook Endoscopy: DPOC-15, Cholangioscopy Access Balloon, G54951, 15 mm, Rx only, Cook Medical, Winston-Salem, NC 27105 This device is intended to guide both optical and accessory devices for diagnostic and therapeutic applications during endoscopic procedures in the pancreaticobiliary system including the hepatic ducts. Also, used for endoscopic removal of biliary stones.

Recall number
Z-0925-2011
Recalling firm
Cook Endoscopy
Product
DPOC-15, Cholangioscopy Access Balloon, G54951, 15 mm, Rx only, Cook Medical, Winston-Salem, NC 27105 This device is intended to guide both optical and accessory devices for diagnostic and therapeutic applications during endoscopic procedures in the pancreaticobiliary system including the hepatic ducts. Also, used for endoscopic removal of biliary stones.
Status
Terminated
Initiated
2010-12-23
Posted
2011-01-19
Terminated
2011-08-04
Reason
Product defect; There is a risk of gastrointestinal perforation due to balloon deflation difficulty or air embolism resulting in poor patient outcomes including death.
Root cause
Other
Action
The firm, Cook Medical, sent an "URGENT: PRODUCT RECALL" letter dated January 7, 2011, to all customers. The letter described the product, problem and the actions to be taken by the customers. The customers were instructed to quarantine any affected product and return to Cook Medical. If they are distributors, they were instructed to contact their customers, direct them to quarantine any affected product and coordinate the return of the affected product to Cook Medical. The customers were also instructed to complete and return the bottom of the letter and send to Cook Medical as well as the PRODUCT RECALL EFFECTIVENESS SURVEY via fax at 812-339-7316 or scanned and emailed to CustomerRelationsNA@cookmedical.com. Note: All non-responding customers will receive a total of three (3) recall notifications ( two (2) follow up requests for a response in addition to the initial recall notification). Should you have any questions, please contact the Customer Relations Department at Cook Medical for more information. Please use CustomerRelationsNA@cookmedical.com or call toll free at 1-800-457-4500, press 4 then enter extension 2146.
Quantity
103 units (total for both units under recall)
Distribution
Worldwide distribution: USA including states of: AL, CA, FL, IL, IN, LA, MA, NC, NY, OH, PA, SC, TN, TX, WA, and WI; and countries of: Belgium, Canada, Germany, Hong Kong, and the Netherlands..
Lot, serial or model codes
Reorder number: DPOC-15, Lot numbers: W2834630, W2835211, W2871208, W2871209, W2889215, W2889216, W2889217, W2889218, W2899398, W2899399, W2902200, W2902201, W2907649, W2907650, W2910395, W2910396, W2914136, W2914137, W2916501, W2916502, W2918945, W2918946, W2920993, W2920994, W2926799, W2926800, W2928562, W2928563, W2930338, W2930882, W2932781, W2932858, W2935202, W2935243 and WR2926438.
510(k) numbers
["K101095"]
PMA numbers
not on file
Product code
FGE
Firm FEI
1037905
Firm city
Winston Salem
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register