Ortho-Clinical Diagnostics: VITROS ECiQ Immunodiagnostic System, Catalog No. 1922814 --- COMMON/USUAL NAME: VITROS ECiQ System --- Ortho-Clinical Diagnostics, 100 Indigo Creek Drive, Rochester, NY 14626 --- All systems that have Software Version 3.8 or below installed and on which the VITROS Total B-hCG II Reagent Pack (human chorionic gonadotropin) is processed. Product Usage: For in vitro diagnostic use only. To perform random access, batch and STAT immunodiagnostic assays on human fluid specimens

Recall number
Z-0924-2012
Recalling firm
Ortho-Clinical Diagnostics
Product
VITROS ECiQ Immunodiagnostic System, Catalog No. 1922814 --- COMMON/USUAL NAME: VITROS ECiQ System --- Ortho-Clinical Diagnostics, 100 Indigo Creek Drive, Rochester, NY 14626 --- All systems that have Software Version 3.8 or below installed and on which the VITROS Total B-hCG II Reagent Pack (human chorionic gonadotropin) is processed. Product Usage: For in vitro diagnostic use only. To perform random access, batch and STAT immunodiagnostic assays on human fluid specimens
Status
Terminated
Initiated
2011-08-24
Posted
2012-01-27
Terminated
2017-05-09
Reason
Software defect; Ortho Clinical Diagnostics received complaints of observed results from diluted samples that were reported as <2.39 mIU/mL (IU/L) instead of the correct No Result, which is accompanied by an Invalid Dilution (ID) code. The diluted results were inconsistent with results from the neat (undiluted) sample, which was within the measuring range of 2.39 - 15,000 mIU/mL. OCD's investigation determined it is possible to obtain negatively biased results if samples are diluted and tested on the VITROS ECi/ECiQ System using VITROS B-hCG II Reagent Packs, Lots 0484 and above.
Root cause
Software design
Action
Ortho Clinical Diagnostics (OCD) sent an URGENT Product Correction Notification letter dated August 24, 2011 via FedEx overnight letter to all US consignees with VITROS ECi/ECiQ Systems. Foreign Affiliates were also notified of the problem on August 24, 2011. The letters provided information regarding the affected product, the problem, the actions to be followed prior to reporting results from their laboratory and a Question and Answer Section. Customers were instructed to 1) Post the enclosed Procedure for Reviewing VITROS Total B-hCG II Results near the VITROS ECi/ECiQ System to advise operators of this issue and assist them in reviewing results from diluted samples; 2) Users should consult with their Laboratory Medical Director and request a physician to resolve any concerns users may have regarding previously reported results from Lots 0484 and above of VITROS Total B-hCG II Reagent Packs; 3) Complete and return the attached Confirmation of Receipt form upon receipt of this notification; and 4) Forward the information in this notification if you have distributed this product outside of your facility. For additional questions contact OCD Customer Technical Services at 1-800-421-331.
Quantity
Domestic: 906, Foreign: 1294
Distribution
Worldwide Distribution - USA (nationwide) and including Puerto Rico and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, and Venezuela.
Lot, serial or model codes
Serial Numbers between 30003000 and 30005177; All VITROS¿ ECi/ECiQ systems that have Software Version 3.8 or below installed and on which the VITROS¿ B-hCG II (human chorionic gonadotropin) Reagent Packs is processed are affected by this correction. --- Note: All systems that are installed worldwide have Software Version 3.8 or lower installed on them, but B-hCG is not processed on all systems worldwide.
510(k) numbers
["K962919"]
PMA numbers
not on file
Product code
KHO
Firm FEI
1000136573
Firm city
Rochester
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register