AGFA Healthcare Corp.: Brand name: AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic Common Name: DR 800

Recall number
Z-0923-2026
Recalling firm
AGFA Healthcare Corp.
Product
Brand name: AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic Common Name: DR 800
Status
Open, Classified
Initiated
2025-11-18
Posted
2025-12-22
Terminated
not on file
Reason
It is possible to exceed the maximum allowed Air Kerma rate (AKR) of 88mGy/min for pulsed fluoro exams. This can happen in case the framerate is increased on the NX workstation when ABS is not enabled.
Root cause
Radiation Control for Health and Safety Act
Action
AGFA NV notified Customers via a letter titled "IMPORTANT PRODUCT INFORMATION LETTER / DR 800", included Product Name, Symptom, Cause, Actions to be taken customer and manufacturers respectively: "Always enable ABS for pulsed fluoro exams. Actions being taken by the manufacturer: Agfa/local business partner will contact you to install a new generator software."
Quantity
35
Distribution
U.S. Nationwide distribution.
Lot, serial or model codes
N/A
510(k) numbers
["K183275"]
PMA numbers
not on file
Product code
IZF
Firm FEI
3001236302
Firm city
Greenville
Firm state
SC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register