Abbott Laboratories, Inc: ARCHITECT c8000 Cuvette Segment, Part #01G46-01/01G06, a component to the ARCHITECT Clinical Chemistry system. There are 11 segments with each cuvette segment holding 15 cuvettes for a total of 165 cuvettes per reaction carousel. The firm name on the label is Abbott Laboratories, Abbott Park, IL. The Abbott ARCHITECT Clinical Chemistry Systems are designed to perform automated chemistry tests utilizing photometry and potentiometric technology. The cuvette segments are a component of the system and are racks that sit in the reaction carousel and hold cuvettes.

Recall number
Z-0922-2018
Recalling firm
Abbott Laboratories, Inc
Product
ARCHITECT c8000 Cuvette Segment, Part #01G46-01/01G06, a component to the ARCHITECT Clinical Chemistry system. There are 11 segments with each cuvette segment holding 15 cuvettes for a total of 165 cuvettes per reaction carousel. The firm name on the label is Abbott Laboratories, Abbott Park, IL. The Abbott ARCHITECT Clinical Chemistry Systems are designed to perform automated chemistry tests utilizing photometry and potentiometric technology. The cuvette segments are a component of the system and are racks that sit in the reaction carousel and hold cuvettes.
Status
Terminated
Initiated
2017-12-08
Posted
2018-01-18
Terminated
2019-07-11
Reason
There is a potential to generate falsely depressed patient results in the cuvettes adjacent to the detached segment post for any assay run on the c4000, c8000, and c16000 instrument.
Root cause
Device Design
Action
The firm, Abbott Laboratories, issued a "Product Correction" letter dated 12/8/2017 via overnight mail on 12/8/2017 to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to adhere to the new instructions listed in Appendix A to avoid damaging the cuvette segments; If a cuvette segment with a detached bottom is identified, replace it before performing any additional testing on your ARCHITECT Clinical Chemistry System; If you have forwarded the product listed above to other laboratories, please inform them of this Product Correction and provide to them a copy of this letter and retain this letter for your laboratory records. The procedures will be updated in a future version of the ARCHITECT Operations Manual. If you or any of the health care providers you serve have any questions regarding this information, U.S. Customers please contact Customer Service at 18774ABBOTT (available 24 hours a day, 7 days a week). Customers outside the U.S., please contact your local area Customer Service.
Quantity
3,357 ARCHITECT c8000 Systems
Distribution
Worldwide Distribution: US (nationwide) including Puerto Rico and countries of: Afghanistan, Albania, Algeria, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Burkina Faso, Cameroon, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Curacao, Cyrus, Czech Republic, Dem Rep of The Congo, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Finland, France, Gaza & Jericho, Georgia, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Ivory Coast, Jamaica, Jordan, Kazakhstan, Kenya, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malawi, Malaysia, Maldives, Mali, Mauritania, Mexico, Moldova, Montenegro, Montserrat, Morocco, Netherlands, New Zealand, Nigeria, North Yemen, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia (CIS), Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, St. Lucia, St. Vincent, Sweden, Switzerland, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad & Tobago, Tunisia, Turkey, UAE, Uganda, United Kingdom, Uruguay, Venezuela, Vietnam, Zambia and Zimbabwe. .
Lot, serial or model codes
All serial numbers
510(k) numbers
["K980367"]
PMA numbers
not on file
Product code
JJE
Firm FEI
1628664
Firm city
Irving
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register