Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Endoscopy: Stryker brand Cart with Isolation Transformer, Model Numbers: 240-099-001, 240-099-001K, 240-099-011, 240-099-011K, 240-099-020, 240-099-020K, 240-099-050, 240-099-072. 240-099-001, 240-099-001K, 240-099-011, 240-099-011K, 240-099-020, 240-099-020K, 240-099-050, 240-099-072, Product is manufactured and distributed by Stryker Endoscopy San Jose, CA The product is indicated for use in the carts include an isolation transfer and a power cord for connecting the transformer to an electrical outlet as means of providing power to the cart for the equipment it houses.

Recall number
Z-0922-2011
Recalling firm
Stryker Endoscopy
Product
Stryker brand Cart with Isolation Transformer, Model Numbers: 240-099-001, 240-099-001K, 240-099-011, 240-099-011K, 240-099-020, 240-099-020K, 240-099-050, 240-099-072. 240-099-001, 240-099-001K, 240-099-011, 240-099-011K, 240-099-020, 240-099-020K, 240-099-050, 240-099-072, Product is manufactured and distributed by Stryker Endoscopy San Jose, CA The product is indicated for use in the carts include an isolation transfer and a power cord for connecting the transformer to an electrical outlet as means of providing power to the cart for the equipment it houses.
Status
Terminated
Initiated
2010-04-16
Posted
2011-01-19
Terminated
2011-01-19
Reason
The electrical cord power plug design may allow the prongs to develop a fracture inside or outside the plug, if the plug is misused.
Root cause
Device Design
Action
Stryker Endoscopy sent an URGENT: Device Recall letter dated April 16, 2010, to Stryker Sales reps and all customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to: Locate the affected units. Visually inspect each cord for damage. If the cord has bent or cracked prongs the unit should be taken out of service immediately. The customer was instructed to call a Regulatory Affairs Recall Representative at (408) 754-2738 for further instruction. If the cord did not show signs of damage, it could remain in service. An acknowledgement form was included to document confirmation of receipt of the letter. Upon completion of the power cord inspection, customers were to sign the acknowledgement form indicating that all of their units were inspected. Customers were asked to mail the form to Stryker Endoscopy or fax to (408) 754-8378 or scan the card and email to PowerCord@stryker.com. For questions regarding this recall call the Stryker RA/QA Department at (408) 754-2738 or (800) 435-0220, Reference RA 2010-034.
Quantity
7371
Distribution
Worldwide Distribution - USA including AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, ID, IL, IN, IA, KS, KY, LA, MA, ME, MI, MN, MS, MO, MT, NE, NH, NJ, NM, NV, MY, NC, ND, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV, and WY and the countries of Canada, Chile, China, India, Korea, Mexico, Singapore, Taiwan and Latin America.
Lot, serial or model codes
Model Numbers/First/Serial Numbers:  240-099-001, 240-099-001K; First Serial number: 070119AB001; 240-099-011, 240-099-011K; First Serial number :070112AD043; 240-099-020, 240-099-020K: First Serial number: 070123AG001 240-099-050; First Serial number: 061206AF001; 240-099-072; First Serial number: 081002AL001.
510(k) numbers
["K935234"]
PMA numbers
not on file
Product code
GCJ
Firm FEI
2936485
Firm city
San Jose
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register