B. Braun Medical, Inc.: B. Braun AccuFlo Elastomeric Infusion Pump 200ML/H 100 ML Generally used for delivery of antibiotic and local anesthesia medications.

Recall number
Z-0921-2014
Recalling firm
B. Braun Medical, Inc.
Product
B. Braun AccuFlo Elastomeric Infusion Pump 200ML/H 100 ML Generally used for delivery of antibiotic and local anesthesia medications.
Status
Terminated
Initiated
2013-12-13
Posted
2014-02-03
Terminated
2014-09-03
Reason
Internal testing of two lots of the AccuFlo CT-2000-100 Elastomeric Infusion Pump identified flow rates which are outside the specified infusion time and may result in overly rapid infusion of medications.
Root cause
Process design
Action
B. Braun Medical sent an Urgent Medical Device Recall letter dated December 13, 2013 to affected customer. The letter identified the product the problem and the action needed to be taken by the customer. The letter was sent to Progressive Medical Inc., the Importer and B. Braun's sole customer. WHAT ACTIONS SHOULD I TAKE? 1. Further use and/or distribution of the affected product should be discontinued immediately. Please identify and quarantine any impacted materials in your possession. 2. Review this recall notification in its entirety and ensure that all users within your organization and any consignees your organization may have are informed of this recall. If you are a distributor, please forward this recall notice to any customers who are in receipt of the impacted product. 3. Determine your current inventory of the affected lots within your facility. Do not destroy any affected product. 4. Utilizing the attached "Recall Acknowledgement" form record the number of full cases and individual units (for any loose units or partial cases) in your possession. If you have no inventory, please indicate that you have zero (0) inventory on the "Recall Acknowledgement" form. 5. Return the completed "Recall Acknowledgement" form to B. Braun Medical Inc. Quality Assurance who will administer the recall on behalf of B. Braun Melsungen AG. The form must be e-mailed to PA_QualityAssurance.BBMUS.Service@bbraun.com or faxed to (610) 849-1197 within two (2) weeks of receipt even it the inventory in your possession is zero. 6. Instructions for Credit, Replacement, Et Return: a.If you have purchased the product from a third-party and have any full cases, partial cases, or unused individual pieces in your possession please contact the seller to obtain instructions for handling these units. b.If you have purchased the product directly from B. Braun Medical Inc. and have any full cases, partial cases, or unused individual pieces of these affected products, ple
Quantity
30, 240 units
Distribution
US Distribution one importer/wholesale distributor in MO.
Lot, serial or model codes
Catalog #CT-2000-100lots 3H0125F151 and 3H0225F151, expiration 7/2018.
510(k) numbers
["K081905"]
PMA numbers
not on file
Product code
MEB
Firm FEI
2523676
Firm city
Allentown
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register