Skip to the content

FDA Medical Device Recalls (openFDA)

Genesys Orthopedic Systems, LLC: 3.75mm x 12mm Fixed Cervical Screws. Components of the Genesys Spine Anterior Cervical Plate System Product Usage: is intended for anterior screw fixation to the cervical spine.

Recall number
Z-0921-2013
Recalling firm
Genesys Orthopedic Systems, LLC
Product
3.75mm x 12mm Fixed Cervical Screws. Components of the Genesys Spine Anterior Cervical Plate System Product Usage: is intended for anterior screw fixation to the cervical spine.
Status
Terminated
Initiated
2013-02-20
Posted
2013-03-08
Terminated
2013-09-09
Reason
Manufacturing residue may be present on the cervical screws. Use of this product may result in patient infection, delayed fusion or non-fusion (pseudarthrosis), inflammatory response, or revision surgery.
Root cause
Employee error
Action
Genesys Orthopedic Systems, LLC sent a Recall Notification letter dated February 20, 2013 to all affected customers. The letter identified the affected product, problem and necessary actions to be taken. Customers were instructed to immediately return the affected product to the Genesys Spine office and complete and return the enclosed Response Form as soon as possible. For questions call 512-381-7070.
Quantity
9 cervical screws
Distribution
USA Nationwide Distribution including the states of: CA and TX.
Lot, serial or model codes
Part Number: GCSF37512 and Lot Number: 0243
510(k) numbers
["K111132"]
PMA numbers
not on file
Product code
KWQ
Firm FEI
3008455034
Firm city
Austin
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register