FDA Medical Device Recalls (openFDA)
Integrated Measurement Systems, Inc.: IMS Patient Scale used on Sunrise Medical's Hoyer Presence and Stature Patient Lifts (Cradle), Models: HOY-4pt-WSC and HOY-StatureWSC
- Recall number
- Z-0921-2007
- Recalling firm
- Integrated Measurement Systems, Inc.
- Product
- IMS Patient Scale used on Sunrise Medical's Hoyer Presence and Stature Patient Lifts (Cradle), Models: HOY-4pt-WSC and HOY-StatureWSC
- Status
- Terminated
- Initiated
- 2007-01-30
- Posted
- 2007-06-19
- Terminated
- 2008-03-02
- Reason
- Patient Scale may become detached, which can cause a patient to fall and result in injury.
- Root cause
- Other
- Action
- The scale manufacturer, IMS, in conjunction with the lift manufacturer, Joerns Healthcare (previously Sunrise Medical), will initiate a field fix strategy consisting of shipping replacement 4-point cradles without scales, for use until a replacement can be designed, tested and validated. Joerns will utilize their Customer Service Help Line (1-800-826-0270) on behalf of IMS, to inform customers of the voluntary field correction, address any questions regarding the IMS scale field fix strategy and to provide additional information on return of affected IMS Patient Scales. Customers were notified by telephone beginning on January 30, 2007.
- Quantity
- 149 units
- Distribution
- Worldwide
- Lot, serial or model codes
- All serial numbers for Model No. HOY-4PTWSC and HOY-STATUREWSC
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- FRW
- Firm FEI
- 3002780774
- Firm city
- Elk Grove Village
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28