FDA Medical Device Recalls (openFDA)
Varian Medical Systems: Flexible Implant tube, blind end, 350 mm, part # GM11003730, labeled in part ***Varian Medical Systems, Inc 3100 Hansen Way Palo Alto, CA 94304 USA www.varian.com***. Available as single unit or contained in the Flexible Interstitial Implant Tube Set, part number GM11004680. For use with the Varian GammaMed Plus brachytherapy afterloader device.
- Recall number
- Z-0921-05
- Recalling firm
- Varian Medical Systems
- Product
- Flexible Implant tube, blind end, 350 mm, part # GM11003730, labeled in part ***Varian Medical Systems, Inc 3100 Hansen Way Palo Alto, CA 94304 USA www.varian.com***. Available as single unit or contained in the Flexible Interstitial Implant Tube Set, part number GM11004680. For use with the Varian GammaMed Plus brachytherapy afterloader device.
- Status
- Terminated
- Initiated
- 2005-04-14
- Posted
- 2005-06-22
- Terminated
- 2005-08-22
- Reason
- A defective treatment applicator in a cancer therapy medical device may become separated potentially causing patient injury.
- Root cause
- Other
- Action
- The recalling firm notified end users by Customer Technical Bulletin on 4/14/05. The bulletin advised of potential for the blind end of the implant tube to become separated with the possible effect of the tube end being inhaled by a patient. Consignees were instructed to stop using product and return for replacement. The firm''s bulletin is also available on the company website.
- Quantity
- 33 devices in domestic distribution
- Distribution
- Product was distributed to direct accounts in CT, MI, PA, CO, VT and MO. Product was additionally sold to foreign accounts in internationally.
- Lot, serial or model codes
- Lot 04/21/1
- 510(k) numbers
- ["K983436"]
- PMA numbers
- not on file
- Product code
- JAQ
- Firm FEI
- 3000206172
- Firm city
- Charlottesville
- Firm state
- VA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28