Signature Orthopedics Pty Limited: Signature Ceramic Femoral Head. Model Number: 111-152-621. The Signature ceramic femoral heads are manufactured by Ceramtec from 72-75% Al2O3+24-26% Z3O3 (Biolox Delta). The heads connect via a 12/14 taper to the stem trunnion, and articulates against the crosslinked UHMWPE liner within the Logical Acetabular Cups. Product Usage: Components of the Signature Orthopaedics hip replacement range are intended to replace a hip joint where bone stock is sufficient to support the implant. When a surgeon has selected prosthetic replacement as the preferred treatment, the devices are indicated for: Non-inflammatory degenerative joint disease including osteoarthritis or avascular necrosis Inflammatory joint disease including rheumatoid arthritis Correction of functional deformity including congenital hip dysplasia Traumatic injury involving the hip joint including traumatic arthritis or femoral head or neck fracture Failed previous hip surgery including internal fixation or joint fusion, reconstruction, hemiarthroplasty, surface replacement, or total replacement

Recall number
Z-0920-2018
Recalling firm
Signature Orthopedics Pty Limited
Product
Signature Ceramic Femoral Head. Model Number: 111-152-621. The Signature ceramic femoral heads are manufactured by Ceramtec from 72-75% Al2O3+24-26% Z3O3 (Biolox Delta). The heads connect via a 12/14 taper to the stem trunnion, and articulates against the crosslinked UHMWPE liner within the Logical Acetabular Cups. Product Usage: Components of the Signature Orthopaedics hip replacement range are intended to replace a hip joint where bone stock is sufficient to support the implant. When a surgeon has selected prosthetic replacement as the preferred treatment, the devices are indicated for: Non-inflammatory degenerative joint disease including osteoarthritis or avascular necrosis Inflammatory joint disease including rheumatoid arthritis Correction of functional deformity including congenital hip dysplasia Traumatic injury involving the hip joint including traumatic arthritis or femoral head or neck fracture Failed previous hip surgery including internal fixation or joint fusion, reconstruction, hemiarthroplasty, surface replacement, or total replacement
Status
Terminated
Initiated
2018-02-02
Posted
not on file
Terminated
2018-08-08
Reason
This recall has been identified due to the batch being mislabeled, with the packaged component's size not corresponding to the label. Therefore, the product intended to be used for the procedure may not be included in the packaging resulting in surgical delay as replacement product is retrieved.
Root cause
Labeling mix-ups
Action
Signature Orthopaedics notified customer(s) of the recall by sending an Urgent Medical Device Recall Letter dated February 2, 2018 via email. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete the Customer Acknowledgement Form and firm's Effectiveness Check Questionnaire.
Quantity
10
Distribution
Worldwide Distribution - U.S. Nationwide and the country of Australia.
Lot, serial or model codes
UDI: 09348215001926. Lot:78560.
510(k) numbers
K121297
PMA numbers
not on file
Product code
MEH
Firm FEI
3008262872
Firm city
Sydney
Firm state
Lane Cove West
Firm country
Australia

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
3e6df46828c1
Size
288213574 bytes
Published
2026-10-06
Read
2026-10-07T03:20:25.099Z
Source file
Source file retained · verify
Refresh due
2026-10-14

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 3e6df46828c1, published 2026-10-06

Search this source