Angiodynamics Worldwide Headquarters: NanoKnife Irreversible Electroporation System. System consists of multiple components: Generator, footswitch, disposable electrode. System is manufactured by AngioDynamics, Inc. --- Catalog #20300101 Intended use: the generator system with six outputs is indicated for surgical ablation of soft tissue.
- Recall number
- Z-0918-2012
- Recalling firm
- Angiodynamics Worldwide Headquarters
- Product
- NanoKnife Irreversible Electroporation System. System consists of multiple components: Generator, footswitch, disposable electrode. System is manufactured by AngioDynamics, Inc. --- Catalog #20300101 Intended use: the generator system with six outputs is indicated for surgical ablation of soft tissue.
- Status
- Terminated
- Initiated
- 2012-01-20
- Posted
- 2012-01-27
- Terminated
- 2016-02-10
- Reason
- Angiodynamics was marketing the NanoKnife System with the ablation zone estimator feature in the US without FDA clearance.
- Root cause
- Labeling False and Misleading
- Action
- AngioDynamics sent an Urgent - Medical Device Correction letter dated December 15, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed not to use the User's Manual that was supplied with the generator when it was purchased. Instructions include discarding their current manual and troubleshooting guide. A replacement manual will be provided with the Field Safety Notice; the troubleshooting guide will not be replaced. Customers were instructed to complete the Reply Form and fax to AngioDynamics at 518-798-1360. If customers further distributed the affected product, they should inform those accounts of the field correction. Angiodynamics will also disable the AZE feature during a schedule service visit. For questions regarding this recall call 1-800-772-6446 or e-mail customerservice@angiodynamics.com.
- Quantity
- 40 units
- Distribution
- Nationwide Distribution
- Lot, serial or model codes
- Batch codes: 00800709, 00880909, 00870909, 00630609, 00610609, 00470109, 00080508, 00180608, 00050508, 00460109, 00830709, 00240708, 00500209, 00930909, 00430109, 00480209, 00371008, 01210911, 00650709, 00670709, 00690709, 00850909, 00920909, 00820709, 00860909, 00890909, 00940909, 00780709, 00770709, 00760709, 00910909, 00790709, 00361008, 00750709, 00510209, 00450109, 00421208, 00490209, 00530209, 00600609
- 510(k) numbers
- ["K080202","K102329"]
- PMA numbers
- not on file
- Product code
- OAB
- Firm FEI
- 3008319439
- Firm city
- Latham
- Firm state
- NY
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28