Philips Medical Systems North America Co. Phillips: Philips OmniDiagnost Eleva with Digital Image (DI) Rel.2 system. The Philips OmniDiagnost Eleva is a diagnostic-imaging device intended for radiographic, fluoroscopic, angiographic, and interventional applications.

Recall number
Z-0918-2011
Recalling firm
Philips Medical Systems North America Co. Phillips
Product
Philips OmniDiagnost Eleva with Digital Image (DI) Rel.2 system. The Philips OmniDiagnost Eleva is a diagnostic-imaging device intended for radiographic, fluoroscopic, angiographic, and interventional applications.
Status
Terminated
Initiated
2009-03-06
Posted
2011-01-18
Terminated
2011-06-16
Reason
Images from two different patients may be combined in one patient folder in the Digital Image (DI) or may be combined as one (printed) image. It is possible that an image of a previous patient registered in the DI will be overwritten with an image of a later patient.
Root cause
Other
Action
On March 6, 2009, Philips sent an "URGENT-Device Correction Easy Diagnost Eleva, MultiDiagnost Eleva, and OmniDiagnost Eleva with DI Rel. 2 systems notification" to their consignees/customers. The consignees/customers are advised that before diagnosing images, always check the stored (Digital Image, Viewform or PACS) or printed images to ensure that these represent the data and images captured from the examination of the respective patient. Do not use images if they do not represent the patient that was examined. Consignees will be contacted by Philips to schedule a software upgrade of their units. The consignees can contact Philips Healthcare Call Center at 800-722-9377, #5, #2 and reference "FCO 70800070" with any questions about this recall.
Quantity
13 units
Distribution
Nationwide distribution: USA including states of: AZ, CA, CO, CT, DE, FL, GA, HI, ID, LA, MA, ME, MI, MN, MO, MS, NC, NJ, NV, NY, OH, OR, PA, SC, TN, TX, VA, WA, WI, and WY.
Lot, serial or model codes
Site Numbers:   533530, 535118, 535257, 535924, 539106, 540860, 541490, 547017, 547142, 547348, 549798, 550258, and 553235.
510(k) numbers
["K032046"]
PMA numbers
not on file
Product code
IZI
Firm FEI
1000524572
Firm city
Bothell
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register